Unofficial package to control ISMATEC products. We are not affiliated with ISMATEC.
Project description
ismatec
Unofficial Python package to control ISMATEC products. We are not affiliated with ISMATEC.
Supported products:
Reglo-CPF Digital
The package is not fully compatible with Windows 10.
Commands that don't work:
set the calibrated flow rate
Getting started
Check the examples folder for example script on how to use the RegloCPF and RegloICC modules.
To connect to the pump in Python, you need to know the comport of the pump when it is connected to the computer. To know this, before plugging the pump into the computer open the Windows Device Manager program and expand Ports. Then plug in the pump and see what new port appears, e.g. COM6. This is what you will use to create the RegloCPF or instance in Python to connect to and control the pump.
Serial settings and protocols
Connection settings according to the manual are
+----------------------+---------------------------------------------+
| Parameter | Comment |
+----------------------+---------------------------------------------+
| Baud rate | 9600 |
+----------------------+---------------------------------------------+
| Parity | None |
+----------------------+---------------------------------------------+
| Handshaking | None |
+----------------------+---------------------------------------------+
| Data bits | 8 |
+----------------------+---------------------------------------------+
| Stop bits | 1 |
+----------------------+---------------------------------------------+
| Address | Between 1 and 8 |
+----------------------+---------------------------------------------+
| Physical connection | RS232 IN (female): 9-pin D-socket on the |
| | the real panel of the RegloCPF, connects |
| | to PC via standard RS232 cable. |
| | RS232 OUT (male): used to connect to other |
| | pumps |
+----------------------+---------------------------------------------+
| Message Terminators | Command string is completed by a carriage |
| | return (ASCII 13, <CR>). The pump confirms |
| | most commands with an asterix *, Yes/No |
| | inquiries are answered by + (yes) or - (no) |
| | Multi-digit replies are concluded by a |
| | carriage return (ASCII 13, <CR>) and a line |
| | feed (ASCII 10, <LF>) |
+----------------------+---------------------------------------------+
Structure of commands:
The address is followed by a character. Some commands have additional parameters that always consist of 4 or 5 figures. The command string is completed by a carriage return (ASCII 13, ). The pump confirms most commands with an asterix *, Yes/No inquiries are answered by + (yes) or - (no). Multi-digit replies are concluded by a carriage return (ASCII 13, ) and a line feed (ASCII 10, ).
Incorrect commands are answered by #. If the pump is in the state of overload, each command is responded with #. Numerical values are confirmed as 3 to 5-digit figures. 4 or 5 digits are numerals, one digit is either a decimal point or a preceding blank space
For a full list of the protocols, check the operating manual: www.ismatec.com/images/pdf/manuals/REGLO_CPF_Digital_new.pdf
Reglo-ICC Digital Peristaltic Pump
Legacy and channel addressing mode are both supported. Event message reading is not supported so event messaging is disabled on instantiation of a RegloICC instance.
There is also a RegloICCFourChannel class that is a subclass of RegloICC with properties to track when the Start, Stop, and Pause channels are called for each pump channel of hte pump. But since Event messaging is not supported, it will not actually reflect the real time state of the pump if it stops pumping on its own.
On testing, the following commands did not work:
set_pump_head_model_type_codeget_actual_calibration_volume- Note: if a pump channel is running and the
Startcommand is used to try to start that channel again, this will cause aCommandErrorto be raised
The following commands have not been implemented:
f- Get current volume/time flow rate (mL/min) -_GET_FLOW_RATE_VOLUME_TIME_MODEf- Set RPM flow rate in volume/time mode (mL/min -_SET_FLOW_RATE_VOLUME_TIME_MODExv- Get time to dispense at a given volume at a given mL/min flow rate. -_GET_TIME_TO_DISPENSE_VOLUME_AT_FLOW_RATExw- Get time to dispense at a given volume at a given RPM -_GET_TIME_TO_DISPENSE_VOLUME_AT_RPMxt- Change factory roller step volume for a particular roller count and tubing size using roller count (6,8,12) -_SET_FACTORY_ROLLER_STEP_VOLUME
Getting started
To connect to the pump in Python, you need to know the comport of the pump when it is connected to the computer. To know this, before plugging the pump into the computer open the Windows Device Manager program and expand Ports. Then plug in the pump and see what new port appears, e.g. COM6. This is what you will use to create the RegloICC instance in Python to connect to and control the pump.
Serial settings and protocols
Connection settings according to the manual are
+----------------------+---------------------------------------------+
| Parameter | Comment |
+----------------------+---------------------------------------------+
| Baud rate | 9600 |
+----------------------+---------------------------------------------+
| Parity | None |
+----------------------+---------------------------------------------+
| Data bits | 8 |
+----------------------+---------------------------------------------+
| Stop bits | 1 |
+----------------------+---------------------------------------------+
| Address | Between 1 and 8 |
+----------------------+---------------------------------------------+
| Physical connection | RS232 IN (female): 9-pin D-socket on the |
| | the real panel of the RegloCPF, connects |
| | to PC via standard RS232 cable. |
| | RS232 OUT (male): used to connect to other |
| | pumps |
+----------------------+---------------------------------------------+
For a full list of the protocols, check the operating manual: http://www.ismatec.com/images/pdf/manuals/14-036_E_ISMATEC_REGLO_ICC_ENGLISH_REV.%20C.pdf
Project details
Release history Release notifications | RSS feed
Download files
Download the file for your platform. If you're not sure which to choose, learn more about installing packages.
Source Distributions
Built Distribution
Filter files by name, interpreter, ABI, and platform.
If you're not sure about the file name format, learn more about wheel file names.
Copy a direct link to the current filters
File details
Details for the file ismatec-1.5.2-py3-none-any.whl.
File metadata
- Download URL: ismatec-1.5.2-py3-none-any.whl
- Upload date:
- Size: 42.6 kB
- Tags: Python 3
- Uploaded using Trusted Publishing? No
- Uploaded via: twine/2.0.0 pkginfo/1.7.0 requests/2.25.1 setuptools/56.0.0 requests-toolbelt/0.9.1 tqdm/4.60.0 CPython/3.9.5
File hashes
| Algorithm | Hash digest | |
|---|---|---|
| SHA256 |
ebf7ae43e56b625c5d9254ff00338882a935649fb8f222a33b58b42645ab382e
|
|
| MD5 |
3a72336e4aa06b0f2f9062e5b3beb66f
|
|
| BLAKE2b-256 |
94a6f5feadd949e5adb66edbaea58d402880bda59709f7645ca1bbcd10d9ee9c
|