JULABO Circulator — MCP Server

Let an AI agent run JULABO heating and refrigerated circulators through JULABO's documented interface commands: read the bath and external temperatures and the heating/cooling power, set the setpoint, start and stop temperature control, decode status and alarm messages, and wait until the bath is stable at temperature.

Package labmcp-julabo
Instruments CORIO CD, CORIO CP, MAGIO MS (and MX), DYNEO DD circulators. PRESTO and older JULABO models use a similar command set but are unverified
Interfaces USB (virtual COM port), RS-232 (CORIO CP, MAGIO, DYNEO)
Protocol JULABO interface commands, appendix "Interface commands" of the original operating manuals: CORIO CD (1.950.0800.us.V10), CORIO CP (1.950.0900.us.V04), MAGIO MS (1.950.1700.us.V05), DYNEO DD (1.950.1300.us.V03). Official downloads: julabo.com
Status 🧪 simulated: tested against a wire-level simulator, not yet verified on hardware. Report a hardware test

Try it without hardware

uvx labmcp-julabo --simulate --check

Connect your circulator

  1. Enable remote control on the circulator. Without it the circulator answers queries but ignores every setting command.
    • CORIO CD/CP: MENU → IntE → rEM → USb (or 232 for RS-232). rOFF appears in standby.
    • MAGIO / DYNEO: Main menu → Connect unit → Remote control → USB or RS232. An "R" appears in the status bar.
  2. Cable: USB-B (the circulator appears as a virtual COM port; on Windows install JULABO's USB driver) or a null-modem RS-232 cable. RS-232 factory settings: 4800 baud, 7 data bits, even parity, 1 stop bit, hardware handshake (RTS/CTS). Interface parameters can only be changed while remote control is off. The server uses these settings but does not enforce the handshake by default (it keeps RTS asserted, which lets the circulator transmit, and USB virtual COM ports often never assert CTS); add ?rtscts=true to the address to enforce it.
  3. Find the port: uvx labmcp ports
  4. Test the connection:
    uvx labmcp-julabo --address /dev/ttyACM0 --check                  # Linux (USB)
    uvx labmcp-julabo --address /dev/tty.usbmodem1101 --check         # macOS
    uvx labmcp-julabo --address COM3 --check                          # Windows
    uvx labmcp-julabo --address "serial://COM3?baudrate=9600" --check # RS-232, non-default baud rate
    
    --check prints the firmware string (version) and the current status, e.g. 02 REMOTE STOP.

Add to your MCP client

Claude Code

claude mcp add julabo -- uvx labmcp-julabo --address /dev/ttyACM0

Claude Desktop / Cursor / Windsurf (claude_desktop_config.json, .cursor/mcp.json, …)

{
  "mcpServers": {
    "julabo": {
      "command": "uvx",
      "args": ["labmcp-julabo", "--address", "/dev/ttyACM0", "--limit", "max_temperature_c=120"]
    }
  }
}

Add --read-only to allow monitoring only (stop_circulation stays available). For other clients: uvx labmcp config julabo-circulator --address /dev/ttyACM0 --client vscode.

Tools

Tool Kind Description
get_command_log 👁 read Return the most recent raw commands sent to / replies received from the instrument (newest last). Useful for debugging and for recording what was done.
get_connection_info 👁 read Report which instrument is connected (identity, address, simulated or real), whether the server is read-only, and the active safety limits. Call this first.
get_setpoint 👁 read Read the current temperature setpoint and whether temperature control is running.
get_status 👁 read Report the circulator's status message (decoded: operating state, rejected command or alarm), whether temperature control is running, the setpoint, the device's own excess-temperature protection setting and warning limits, and the firmware version.
read_temperatures 👁 read Read the bath temperature, the heating/cooling power in % and the safety-sensor temperature, and optionally the external Pt100 sensor.
reconnect 🛑 safety Close and re-open the connection to the instrument (e.g. after it was power cycled or a cable was re-plugged).
set_setpoint ⚠️ hazard Set the circulator's temperature setpoint (out_sp_00). If temperature control is running, the circulator starts heating or cooling to it immediately. Refused outside min_temperature_c..max_temperature_c; the circulator itself also rejects values outside its range or warning limits (reported as an error). Needs remote control enabled.
start_circulation ⚠️ hazard Start temperature control (out_mode_05 1): the pump runs and the bath heats or cools to the setpoint, which is checked against the safety limits first (it may have been changed on the front panel). Make sure the bath is filled and any external hoses are connected and secured. Needs remote control enabled.
stop_circulation 🛑 safety Stop temperature control and the pump (out_mode_05 0), then confirm with in_mode_05. Also ends a running wait_for_temperature. The bath stays hot/cold after stopping.
wait_for_temperature 👁 read Poll the bath (or external) temperature until it has been within tolerance_c of the target for stable_for_s seconds, or timeout_s passes. Does not change anything; start circulation first. Stops early (reached = false) if the circulator raises an alarm or a stop command is sent. Returns a short temperature trace.

Safety limits

Limit Default Meaning
max_temperature_c 90 °C Highest setpoint an agent may set (also checked before start_circulation)
min_temperature_c 5 °C Lowest setpoint an agent may set (also checked before start_circulation)
max_wait_s 7200 s Longest wait_for_temperature

The defaults suit a water bath. Set them to your bath fluid's working range, e.g. --limit min_temperature_c=-30 --limit max_temperature_c=150 for a silicone oil / glycol-rated setup. The circulator additionally enforces its own working range, warning limits and excess-temperature protection; rejections come back as errors such as -11 VALUE TOO LARGE.

Example prompts

  • "Set the circulator to 37 °C, start it and tell me when the bath has been stable within ±0.1 °C for five minutes."
  • "What's the bath temperature and how hard is the circulator working right now?"
  • "Cool the jacketed reactor to 10 °C using the external Pt100 as the reference and let me know when it gets there."
  • "Is there any alarm on the circulator? Explain what it means."
  • "Stop the circulator."

Notes

  • Remote control is a front-panel setting. In manual mode (00 MANUAL STOP / 01 MANUAL START) the server refuses to send settings and tells you how to enable remote control.
  • Every setting is verified. OUT commands have no reply, so after each one the server queries status (to catch -08 INVALID COMMAND, -09 COMMAND NOT ALLOWED IN CURRENT OPERATING MODE, -10 VALUE TOO SMALL, -11 VALUE TOO LARGE, -13 VALUE EXCEEDS TEMPERATURE LIMITS) and reads the setpoint or in_mode_05 back. stop_circulation retries once and reports an error if the circulator still says it is running.
  • Model differences. CORIO CD has only the basic set (actual value, power, safety sensor, setpoint, start/stop). Warning limits (in_sp_03/04) need CORIO CP, MAGIO or DYNEO; the external Pt100 (in_pv_02) needs MAGIO or DYNEO. The server detects missing commands (no reply) and reports those values as null.
  • Command pacing (unverified). The manuals give no minimum time between commands. The server waits command_delay_s (default 0.25 s) after each setting command before checking status; change it with --option command_delay_s=0.5 if your circulator misses commands.
  • Watchdog. MAGIO/DYNEO circulators have a watchdog configured in their own menu (mode, timeout, fallback setpoint, reset on "setpoint only" or "all valid commands"). The server does not configure it. With restart mode "all valid commands" you can keep it fed while the server runs with --option keepalive_s=10 (queries status every 10 s). JULABO's alarm list notes that it expects the setpoint at least every 30 s in "setpoint" restart mode, which this keep-alive does not do.
  • Status and alarm texts are decoded from the code (-01, -14, -63, ...) using the alarm tables in the manuals; the exact wording the circulator sends is passed through unchanged.
  • Not implemented: pump stage/capacity (out_sp_07/27), warning-limit writes (out_sp_03/04), controller parameters (out_par_*), external/internal control switch (out_mode_04), EPROG and ATC calibration commands.

Hardware verification

Model Firmware Interface Verified by Date
none yet: be the first

Release files for labmcp-julabo 0.1.0

For a detailed explanation of source distributions (sdists) and built distributions (wheels), please see the package formats documentation.

Source distribution (sdist)

Source distribution for labmcp-julabo 0.1.0
File Size Uploaded
labmcp_julabo-0.1.0.tar.gz 16.9 kB Details

Built distribution (wheel)

Table of built distributions (wheels) for labmcp-julabo 0.1.0
File Interpreter ABI Platform
labmcp_julabo-0.1.0-py3-none-any.whl Python 3 none any Details

Total release size: 34.0 kB

Release files / labmcp_julabo-0.1.0.tar.gz

Download URL labmcp_julabo-0.1.0.tar.gz
Size 16.9 kB
Tags Source
SHA-256 checksum
How to use checksums
b3292fcad4dbe2bbe0311dddf24151393f8ec351c41e520cf315545fef0d2ec4
BLAKE2b-256 checksum
How to use checksums
f43e8868b6a33939806a39ee232824b2af450e01786d5d9deee5878ae2d8ff63
Upload date
Uploaded using Trusted Publishing?
What is trusted publishing?
Yes
Uploaded via twine/7.0.0 CPython/3.13.14

Provenance

Provenance describes where a file came from. On PyPI, provenance is shared via attestations, which provide a verifiable record of the build or publishing details. View details, limitations and caveats.

PyPI Publish Attestation

PyPI verified that this artifact, at this checksum, originated from the publisher listed below.

Signed by GitHub Actions, verified by PyPI on Sep 26, 2026.

Transparency log

Release files / labmcp_julabo-0.1.0-py3-none-any.whl

Download URL labmcp_julabo-0.1.0-py3-none-any.whl
Size 17.1 kB
Tags Python 3
SHA-256 checksum
How to use checksums
f3ae61dff9bec990830ada216a9133e18b490971075fa07c45afe09e2e623891
BLAKE2b-256 checksum
How to use checksums
f7bde3689cdc216d6e087eaee7c9a335f96f14a9ab45840f6778f26e0c0761b2
Upload date
Uploaded using Trusted Publishing?
What is trusted publishing?
Yes
Uploaded via twine/7.0.0 CPython/3.13.14

Provenance

Provenance describes where a file came from. On PyPI, provenance is shared via attestations, which provide a verifiable record of the build or publishing details. View details, limitations and caveats.

PyPI Publish Attestation

PyPI verified that this artifact, at this checksum, originated from the publisher listed below.

Signed by GitHub Actions, verified by PyPI on Sep 26, 2026.

Transparency log

Release history Release notifications | RSS feed

0.1.2

2 release files

0.1.1

2 release files

This release

0.1.0 This release

2 release files

Anthropic, PBC Visionary sponsor Bloomberg Visionary sponsor Hudson River Trading Visionary sponsor Meta Visionary sponsor NVIDIA Visionary sponsor Microsoft Sustainability sponsor Depot Continuous Integration AWS Cloud computing and Security Sponsor Datadog Monitoring Fastly CDN Google Download Analytics Sentry Error logging StatusPage Status page