open289

English | 简体中文

CI

An open Python package, command-line interface and MCP server for the Fluke 287/289 multimeter, read over a Fluke IR3000 FC infrared-to-Bluetooth-LE adapter, written so that an AI agent (Claude Code, or any MCP client) can read, log and download real measurements by itself, without Fluke Connect. It reads the live value (about 4 readings/s), the full display state (function, range, HOLD / MIN MAX / REL modes, MIN / MAX / AVERAGE), everything saved in the meter (measurements, MIN MAX and peak sessions, trend recordings), logs to CSV or JSONL for hours with automatic reconnect, downloads trend recordings, and can press the meter's front-panel keys.

The meter is reached through the adapter's pass-through to the meter's own serial command set (ID, QM, QDDB, QSRR, ...); the protocol was worked out on real hardware, see docs/PROTOCOL.md.

Status and tested hardware

Version 0.1.1 (beta). What was verified on real hardware on 2026-11-06: an IR3000 FC with radio firmware 01.00.01 and MSP firmware 01.00.15, a Fluke 289 with firmware V1.71, a Windows 11 PC with Intel Bluetooth (Python and bleak over WinRT).

  • Only a 289 was tested. The 287 speaks the same protocol and should work; that is untested.
  • Linux and macOS: bleak supports them and the package has no Windows-only code in the Bluetooth path, but real hardware was never tried there. CI runs the whole test suite on Linux and Windows against the virtual meter only (see Virtual meter).
  • Measured speeds: live readings arrive about 4 times per second; one pass-through command takes about 1.3 s (QDDB display snapshots ran about 1.2 to 1.8 s apart); a trend recording downloads at 0.44 s per record (up to 8 records per round-trip, as many as fit in 87 bytes).

Install

Requirements: Python 3.10+, a Bluetooth LE radio, a Fluke 287/289 and a Fluke IR3000 FC.

pip install open289

This installs the commands open289 and open289-mcp (python -m open289 ... is the same as open289 ...) and the single dependency bleak.

From a clone, with uv or a plain venv:

uv venv --python 3.12 .venv
uv pip install --python .venv\Scripts\python.exe -e .
python -m venv .venv
.venv\Scripts\pip install -e ".[test]"        # Linux / macOS: .venv/bin/pip

Windows notes:

  • Switch Bluetooth on (Settings, Bluetooth and devices). No pairing or bonding is needed.
  • The console may use a legacy code page (GBK, CP1252); measurement strings contain µ, Ω and °. The CLI reconfigures its own stdout and stderr to UTF-8, and the launchers in a clone start Python with -X utf8 for everything else: open289.cmd live -n 10 (Windows) and ./open289.sh live -n 10 (POSIX shell, Git Bash). With a pip install, python -X utf8 -m open289 ... does the same.
  • If pip's Scripts folder is not on PATH, call python -m open289 or give the full path to open289.exe.

Hardware setup

  1. Clip the IR3000 FC onto the meter's infrared port on the back of the 287/289 (the meter must be switched on).
  2. Wake the adapter: press its button for more than 1 s until the LED flashes green. It advertises over Bluetooth only while it is awake and not connected.
  3. Close Fluke Connect on any phone or tablet. The adapter accepts exactly one Bluetooth connection. While another program holds it, open289 reports NOT_FOUND (the adapter does not advertise) or BUSY. open289 itself takes a lock, so two open289 processes never fight over the adapter: the second gets BUSY.
  4. Check: open289 scan lists the adapter; open289 info shows the meter. info reports "meter": {"attached": false} when the adapter is awake but not on a powered meter.

Adapter and meter both go to sleep or off by themselves and cannot be switched on remotely; wake the adapter with its button and the meter with its dial.

Quick start

open289 scan                                # find the adapter
open289 info                                # adapter and meter identity, batteries, clock, saved counts
open289 read                                # the whole display now: function, range, modes, every reading
open289 live -n 20                          # 20 real-time readings, one JSON line each
open289 log --duration 10m --out volts.csv  # log to a file; ends by itself, or: open289 stop / Ctrl+C
open289 stop                                # end a running live / read --every / log cleanly

Every command prints JSON (one object per line) unless --text is given; add --sim to try everything without hardware. Output of the virtual meter (never invented; hardware values differ, the fields are the same):

$ open289 --sim scan
{"ok":true,"cmd":"scan","devices":[{"address":"FA:28:90:00:00:01","name":"IR 3000 FC","rssi":-60}]}

$ open289 --sim info
{"ok":true,"cmd":"info","adapter":{"address":"FA:28:90:00:00:01","name":"IR 3000 FC","radio_fw":"01.00.01","msp_fw":"01.00.15","battery_pct":100},"meter":{"attached":true,"model":"FLUKE 289","firmware":"V1.71","serial":"12345678","battery":"FULL","memory_level":"OK","clock_epoch":1794013461,"clock":"2026-11-07 01:04:21.000"},"saved":{"recordings":1,"minmax":0,"peak":1,"measurements":2}}

$ open289 --sim --text info
adapter : FA:28:90:00:00:01  radio fw 01.00.01  MSP fw 01.00.15  battery 100%
meter   : FLUKE 289  firmware V1.71  serial 12345678
battery : FULL   memory: OK   clock: 2026-11-07 01:04:31.000 (meter local time)
saved   : {'recordings': 1, 'minmax': 0, 'peak': 1, 'measurements': 2}

info.meter.clock is the meter's own clock (local time, as set on the meter), not the PC's.

Real-time acquisition

Command Source Rate Carries
live the adapter's BLE notifications ~4/s, one JSON line per notification the primary reading and, when the meter shows one, the secondary reading
read one QDDB display query through the pass-through one round-trip, ~1.3 s the whole display: function, secondary function, range, autorange, modes, every reading
read --every 1s the same, repeated about one per interval (at least ~1.3 s) one snapshot per line; ends with a done line

Use live for fast values and logging, read when you need to know what the meter is doing (HOLD? MIN MAX? REL? which range?), and read --every when you need modes or MIN / MAX / AVERAGE over time. live and read can be combined by log --snapshot-every (next section) on one connection.

$ open289 --sim live -n 3
{"type":"reading","t":1794027861.982,"value":5.0625,"unit":"V DC","display":"5.0625 V DC","state":"NORMAL","function":"VDC"}
{"type":"reading","t":1794027862.227,"value":5.0625,"unit":"V DC","display":"5.0625 V DC","state":"NORMAL","function":"VDC"}
{"type":"reading","t":1794027862.481,"value":5.0625,"unit":"V DC","display":"5.0625 V DC","state":"NORMAL","function":"VDC"}
{"ok":true,"cmd":"live","event":"done","readings":3,"duration_s":0.748,"rate_hz":4.01,"stopped_by":"count"}

A live reading line

Field Meaning
type "reading"
t PC time of the notification, Unix seconds (3 decimals)
value the number in SI base units (V, A, ohm, F, Hz, s, S; degrees C / F and percent as they are): 1.5 mV is 0.0015. null for an overload (OL) or a blank / invalid display
unit the base unit (without the SI prefix; display carries the prefix): V DC, V AC, V AC+DC, A DC, Ω, S, Hz, F, s, °C, °F, %, dB, dBV, dBm, CF
display the text the meter shows, with the SI prefix and decimals of the display (5.0625 V DC, OL V AC)
state NORMAL, OL_RANGE (overload), BLANK (autorange blanking), INVALID (during a dial change)
function the live function name (VDC, VAC, mVDC, Ω, Continuity, Diode, Temp, mADC, ...)
attribute present when the function has one: positive edge for Hz, continuity open / short / glitch, a good diode
capture true when the reading is a captured value (HOLD or similar)
secondary present when the meter shows a secondary value: an object with the same value, unit, display, state, function fields
raw with --raw only: the 16 notification bytes in hex

The primary and secondary readings are what the meter's main and secondary display show. With Hz on a voltage function the Hz value becomes primary and the volts the secondary, as on the meter:

$ open289 --sim live -n 1            # V AC with the Hz secondary function
{"type":"reading","t":1794027906.375,"value":60.002,"unit":"Hz","display":"60.002 Hz  [positive edge]","state":"NORMAL","function":"VAC","attribute":"positive edge","secondary":{"value":null,"unit":"V AC","display":"OL V AC","state":"OL_RANGE","function":"VAC"}}

What live does not carry: no mode flags (HOLD, MIN MAX, REL, RECORD), no MIN / MAX / AVERAGE values, no range, no function name in the display-snapshot vocabulary. In HOLD the primary is the held value, in REL the relative value. Use read (or log --snapshot-every) for those. A live run ends at -n, --for or open289 stop and prints a done line (readings, duration_s, rate_hz, stopped_by: count, duration, stop); Ctrl+C ends it without a summary (exit 130). --timeout (default 10 s) turns a silent link into an error instead of waiting forever.

A display snapshot (read)

$ open289 --sim read
{"ok":true,"cmd":"read","t":1794027861.456,"function":"V_DC","secondary_function":"NONE","autorange":true,"range":5.0,"unit":"VDC","modes":["HOLD"],"readings":[{"id":"LIVE","value":5.0613,"unit":"VDC","display":"5.0613 V DC","state":"NORMAL","attribute":"NONE","decimals":4,"time":"2026-11-07 01:04:21.456","ts":1794013461.4560547},{"id":"PRIMARY","value":5.0625,"unit":"VDC","display":"5.0625 V DC","state":"NORMAL","attribute":"NONE","decimals":4,"time":"2026-11-07 01:03:58.325","ts":1794013438.3251953}],"hazard":false,"primary":{"id":"PRIMARY",...}}
Field Meaning
t PC time of the query, Unix seconds
function the primary function: V_DC, V_AC, MV_DC, OHMS, CONTINUITY, DIODE_TEST, CAPACITANCE, TEMPERATURE, MA_DC, ...
secondary_function NONE, HERTZ, PEAK_MIN_MAX, ...
autorange, range autorange on / off and the range's full-scale value (in base units; 5.0 = the 5 V range)
unit the primary unit code (VDC, OHM, CEL, ...)
modes active modes, e.g. HOLD, MIN_MAX_AVG, REL, RECORD; empty list = none
readings every number on the screen, in display order (below)
primary a copy of the PRIMARY entry of readings
hazard true when the meter shows its high-voltage warning (bolt)
minmax_start when MIN MAX was started, local meter time (only when MIN MAX is active)

Each readings entry has: id (LIVE, PRIMARY, SECONDARY, MINIMUM, MAXIMUM, AVERAGE, REL_REFERENCE, TEMP_OFFSET, ...), value (SI base units, null for OL), unit, display, state, attribute, decimals, time (meter local time) and ts (the same meter clock value as a number: local wall-clock time encoded as epoch seconds, no timezone; not comparable with t). LIVE is the instantaneous input; PRIMARY is the main reading, which differs from LIVE in HOLD or REL. With MIN MAX active the snapshot also has MINIMUM, MAXIMUM and AVERAGE:

$ open289 --sim press MINMAX ; open289 --sim --text read
00:05:09.566  V_DC   range AUTO 5   modes MIN_MAX_AVG
    LIVE           5.0626 V DC            2026-11-07 01:05:09.567
    PRIMARY        5.0626 V DC            2026-11-07 01:05:09.567
    MINIMUM        5.0612 V DC            2026-11-07 01:05:09.099
    MAXIMUM        5.0631 V DC            2026-11-07 01:05:07.500
    AVERAGE        5.0621 V DC            2026-11-07 01:05:09.567

read --every 1s [-n N] [--for 5m] prints one snapshot per line (ok, cmd, t, ...), an {"ok":false,...} line when one query fails (for example METER_BUSY) without ending the run, and a final {"event":"done","snapshots":N, "stopped_by":...}.

Logging

log writes live readings to a file, in the foreground, until --duration, open289 stop or Ctrl+C. It never stops because of a lost link as long as --reconnect-for allows retries.

open289 log --duration 10m --out volts.csv
open289 log --duration forever --out run.jsonl --snapshot-every 10s     # until: open289 stop
open289 log --every 1s --duration 8h --out night.csv                    # one reading per second
Option Default Meaning
--out, -o data/log_<time>.csv output file (.csv or .jsonl; relative to the current folder; the default is in the data folder)
--format csv|jsonl from the --out extension, else csv file format
--duration, -d forever logging time, counted from the first connection: 30s, 10m, 8h, 1.5h, 2d, forever
--every 0 one reading per interval (the most recent reading at each tick); 0 = every notification, ~4/s
--snapshot-every 0 (off) also record the full display (modes, MIN / MAX / AVG) at this interval, e.g. 10s
--status-every 10s status line on stdout; 0 = off
--reconnect-for 5m keep retrying this long after the link is lost; 0 = end the log at the first loss

CSV columns: pc_time (Unix seconds), local_time (PC local time with milliseconds), value, unit, display, state, function, attribute, secondary_value, secondary_unit, secondary_display:

pc_time,local_time,value,unit,display,state,function,attribute,secondary_value,secondary_unit,secondary_display
1794027865.921,2026-11-07 00:04:25.921,5.062,V DC,5.0620 V DC,NORMAL,VDC,,,,
1794027866.168,2026-11-07 00:04:26.167,5.062,V DC,5.0620 V DC,NORMAL,VDC,,,,

value is in SI base units and empty for an overload. With --snapshot-every, a CSV log puts the snapshots in a sidecar file <file>.display.jsonl, one JSON object per line: t (Unix seconds), ack, hex (the raw QDDB answer) and display (the same object as a read snapshot). A failed snapshot query is written as {"t":..., "ack":"ERROR", "error":{...}} and the log goes on.

JSONL has one object per line, with a type:

type Content
ble a reading: the fields of a live line (t, value, unit, display, state, function, attribute, secondary) plus hex, the 16 notification bytes
qddb a display snapshot (t, ack, hex, display), only with --snapshot-every
event link_lost (with reason) and reconnected, with t
{"t": 1794027869.307, "value": 5.0612, "unit": "V DC", "display": "5.0612 V DC", "state": "NORMAL", "function": "VDC", "type": "ble", "hex": "b4c50008020c00000000000000000000"}

Decimation. --every 1s writes one row per second: the latest reading at that moment. The statistics in the summary then cover only the written rows. Use --every 0 when you need every notification (or to see transients), --every for long runs.

Status lines on stdout while it runs (a started line first, then connected once the adapter is found and armed, then a status line every --status-every):

{"ok":true,"cmd":"log","event":"started","file":"...x.csv","format":"csv","duration_s":3.0,"every_s":0.0,"snapshot_every_s":1.0}
{"ok":true,"cmd":"log","event":"connected","connect_s":2.41,"model":"FLUKE 289","adapter":"FA:28:90:00:00:01"}
{"ok":true,"cmd":"log","event":"status","elapsed_s":3.5,"records":4,"last":"5.0627 V DC"}

Connecting takes time. Every command first has to see the adapter advertise. The IR3000 FC advertises only every few seconds, and right after a previous connection ends it can take about 20 s before it advertises again (measured on hardware). --duration therefore counts from the first connection (connect_s says how long finding the adapter took), so a --duration 30s log really holds about 30 s of readings.

Reconnect and gaps. When the Bluetooth link drops (meter off, adapter asleep, out of range, "no live reading for 10 s") the log prints link_lost (with retry_for_s), writes a link_lost event into a JSONL file, retries every 3 s, and on success prints reconnected (with gap_s) and carries on appending to the same file. If the link does not come back within --reconnect-for, the log ends with stopped_by: "link_lost" and the data written so far is kept. It fails with LINK_LOST only when not one reading was logged; a log that never connected at all fails at once with the real error (NOT_FOUND, DETACHED, ...).

Stopping. open289 stop (from another terminal or process) ends it within about a second; so do Ctrl+C, the duration, and an unrecoverable link loss. The file is flushed after every row, so a killed process loses at most the last row.

Summary (the last line, "event":"done"):

{"ok":true,"cmd":"log","event":"done","file":"...x.csv","format":"csv","records":12,"packets":12,"snapshots":2,"started":1794027863.2,"ended":1794027868.7,"duration_s":5.535,"connect_s":2.41,"logging_s":3.125,"stopped_by":"duration","gaps":[],"stats":{"V DC":{"n":12,"min":5.0614,"max":5.0629,"mean":5.0622}},"meter":{"model":"FLUKE 289","adapter":"FA:28:90:00:00:01"},"display_file":"...x.csv.display.jsonl"}

records rows written, packets notifications received, snapshots display snapshots taken, duration_s the whole command, connect_s the time until the first connection, logging_s the time from then on, stopped_by (duration, stop, link_lost, interrupted), gaps (a list of {from, to, reason}, Unix seconds), stats (n, min, max, mean of the primary value per unit; overloads are left out) and display_file (CSV with snapshots only). --text prints a two-line version with the same numbers.

Saved data

open289 memory                       # everything saved in the meter (--text --detail lists every reading)
open289 recording 0 --out rec0.csv   # download trend recording 0
open289 recording 0 --out rec0.csv --start 40   # resume an interrupted download (appends)

memory reads the counts and then every saved item, and prints one JSON object: counts (measurements, minmax, peak, recordings) and items. Each item has kind (measurement, minmax, peak, recording), index (the index within its kind: this is what recording INDEX takes), name, seq, the same fields as a display snapshot (function, modes, readings, primary, ...), start / end for sessions, and for recordings interval_s, event_threshold, records, reading_index and duration_s. If some items cannot be read the result has "ok":false and an errors list, and the exit code is 6 (the readable items are still listed). Every saved item is one query, so memory has a 600 s limit instead of 120 s.

$ open289 --sim --text memory
saved items: {'recordings': 1, 'minmax': 0, 'peak': 1, 'measurements': 2}
measurement #0   'Save 1'         2026-09-29 03:15:52.785  TEMPERATURE  34.7 °C
measurement #1   'Save 2'         2026-11-01 01:56:42.182  CONTINUITY  OL Ω [OPEN_CIRCUIT]
peak        #0   'Save 3'         2026-11-06 21:29:01.471  V_DC  0.005 V DC
recording   #0   'Save 4'         2026-11-06 21:40:19.248  V_DC  1.9874 V DC  72 records, interval 1 s, 1.0 min

Recording download. recording INDEX fetches the trend records of one recording (the meter's RECORD mode) and writes one row per record. It batches up to 8 queries per round-trip and takes 0.44 s per record on the hardware (--no-batch: one per round-trip, about 1.4 s). A record that fails is retried (up to 3 more times). If the meter shows a menu, the download stops with METER_BUSY and a hint with the record to resume from; so does a failing record (PROTOCOL_ERROR). Resume with the same --out and --start K: the file is appended to, without a second header. --limit K stops before record K (an end index, not a count). Events: started, progress every 10 s (done, of), and done (records, duration_s, s_per_record, retries, file).

CSV columns (JSONL has the same keys): index, start, end (meter local time), unit, primary, primary_state, minimum, minimum_state, average, average_state, maximum, maximum_state, samples, record_type (INTERVAL, ...), stable, transient. Values are in SI base units; an _state of OL means the value is empty.

index,start,end,unit,primary,primary_state,minimum,minimum_state,average,average_state,maximum,maximum_state,samples,record_type,stable,transient
0,2026-11-06 21:40:19.248,2026-11-06 21:40:19.752,VDC,1.4904,NORMAL,1.4903,NORMAL,1.49032,NORMAL,1.4904,NORMAL,5,INTERVAL,STABLE,NON_T

Front-panel keys

press KEY [KEY ...] simulates the keys in order and returns the display state after each one. Keys: F1 F2 F3 F4 UP DOWN LEFT RIGHT HOLD MINMAX RANGE INFO BACKLIGHT ONOFF.

open289 press HOLD                    # toggle HOLD
open289 press RANGE                   # step the range (manual)
open289 press RANGE --held            # back to autorange
open289 press MINMAX                  # start MIN MAX
open289 press MINMAX F4               # stop it (opens the result page), then close the page
open289 press ONOFF F1 --allow-off    # switch the meter off for good

Options: --held sends a long press (HELD_DOWN) for every key; --settle S (default 1.0) waits that long before the state is read back; --allow-off permits ONOFF. Each entry of the result has key, held, ack (OK) and state (a display snapshot; or a text when the meter shows a page, see below). press stops at the first key the meter does not acknowledge. press is the only command that changes the meter's state (apart from raw --unsafe).

$ open289 --sim press HOLD
{"ok":true,"cmd":"press","presses":[{"key":"HOLD","held":false,"ack":"OK","state":{"function":"V_DC","secondary_function":"NONE","autorange":true,"range":5.0,"unit":"VDC","modes":[],...}}]}

What was observed on the hardware:

  • HELD_DOWN is a long press. RANGE held = back to autorange. Held MINMAX behaves like a press.
  • MIN MAX has pages. The first MINMAX starts MIN MAX (modes has MIN_MAX_AVG). The second stops it and opens the result page (F1 restart, F2 save, MINMAX or F4 close). F2 on the result page opens the save page (F1 adds to the name, F2 saves, F4 closes). Closing a page returns to the measurement screen.
  • Why queries return METER_BUSY there. While a menu, result page, save page or setup screen is open, the meter answers every measurement query (QDDB, QSLS, ...) with an execution error. open289 names it METER_BUSY (exit 5). press then shows "state": "menu / result page open (measurement queries blocked)". Close the page with the keys (F4) or on the meter, then retry.
  • BACKLIGHT cycles off, bright, dim, also on the pages.
  • ONOFF is refused (UNSAFE_OPERATION, exit 7) unless --allow-off. With it, the meter asks for confirmation (turning off meter will break communication): F1 confirms, F4 cancels. After F1 the meter is off, the adapter switches itself off a moment later, and neither can be switched on remotely. This cannot be undone from the PC.

Raw commands

raw COMMAND [COMMAND ...] sends meter commands through the pass-through and returns each answer (ack, and text or, for binary replies, binary_hex). Query-only by default: ID, IM and Q* commands (QM, QDDB, QDDA, QBL, QSLS, QMP clock, ...) pass; anything else (MP, CSD = clear memory, RI = reset, OFF, PRESS, ...) is refused with UNSAFE_OPERATION before it is sent. --unsafe lifts the filter; use it only when you know exactly what the command does, because some commands erase saved data or switch the meter off.

$ open289 --sim raw ID QM
{"ok":true,"cmd":"raw","results":[{"command":"ID","ack":"OK","text":"FLUKE 289,V1.71,12345678"},{"command":"QM","ack":"OK","text":"5.0628E0,VDC,NORMAL,NONE"}]}

$ open289 --sim raw MP
{"ok":false,"cmd":"raw","results":[{"command":"MP","error":{"code":"UNSAFE_OPERATION","message":"meter command b'MP' is not a read-only query","hint":"only ID, IM and Q* queries pass by default; add --unsafe for other commands"}}]}

Each command is sent and answered in turn; a refused or failed one is reported in its own result and the exit code is that of the first error (7 for a refusal). A command over 87 bytes is refused before it is sent (see Safety).

Self-test

open289 selftest is a read-only health check of the adapter, the link and the meter (it never changes the meter):

Check What
S01 adapter advertising
S02 connect and arm the adapter control point
S03 meter attached (FLUKE 287/289, not the detached FLUKE PC3000)
S04 adapter firmware answers
S05 pass-through ID
S06 display snapshot (QDDB) decodes (the meter must be on its measurement screen)
S07 a live reading arrives within 5 s
S08 adapter battery >= 20 %
S09 meter battery not EMPTY / ALMOST_EMPTY

The result lists every check with pass, measured and expected, plus passed and failed counts; the exit code is 1 when a check failed. A failed S01, S02 or S03 ends the run early (the later checks need that link).

$ open289 --sim selftest
{"ok":true,"cmd":"selftest","passed":9,"failed":0,"items":[{"id":"S01","name":"adapter advertising","pass":true,"measured":"FA:28:90:00:00:01","expected":null},...]}

Virtual meter

The package contains a model of an IR3000 FC with a Fluke 289 attached (open289/sim.py), so the whole tool, the MCP server and the test suite run with no Bluetooth and no hardware. Select it with --sim or OPEN289_SIM=1.

It simulates what was measured on the real hardware: the GATT table and the CNX ID write that arms the adapter, the 18-byte control-point write limit, the pass-through framing, the out-buffer overflow (no reply above 128 bytes, a modelled guess; 181 measured) and the adapter hang at 373 bytes and more (open289 itself refuses anything over 87 bytes) when it is overrun, the binary QDDB / QSMR / QMMSI / QPSI / QRSI / QSRR layouts, the 16-byte live notifications at 4/s, the single connection (a second process is refused and the adapter stops advertising), the front-panel keys with their pages (MIN MAX result and save pages, power-off confirmation, backlight cycle), a meter that is switched off and an adapter that goes to sleep, saved memory with two measurements, a peak session and a 72-record recording. Behaviour that was never observed on hardware is marked GUESS in the code. Identity: address FA:28:90:00:00:01, serial 12345678. It is a stand-in for tests and demos, not a measurement instrument: the "readings" are synthetic.

The state of the virtual device is one JSON file shared by every process (state/sim_state.json in the data folder), so open289 press HOLD in one command is seen by the next.

python -m open289.sim show                            # the state as JSON
python -m open289.sim reset                           # factory state
python -m open289.sim power-cycle                     # switch adapter and meter on again (clears an adapter hang)
python -m open289.sim set function=V_AC input=open meter_on=false
python -m open289.sim audit [--clear]                 # every GATT write the tool made (state/sim_writes.jsonl)

set keys: function (V_DC V_AC MV_DC MV_AC OHMS CONTINUITY DIODE_TEST TEMPERATURE MA_DC, and V_AC_HZ / V_DC_HZ for the Hz secondary function), secondary (NONE / HERTZ), input (the input of the current function; open for OL), inputs.<FUNCTION>, hz, meter_on, adapter_on, adapter_hung, autorange, range_index, screen (normal, minmax_result, save_page, power_off_confirm), backlight, modes, seed, clock_offset. Environment: OPEN289_SIM_RATE (live notifications per second, default 4), OPEN289_SIM_DELAY (latency of a pass-through send, default 0.05 s; the hardware takes about 0.6 s), OPEN289_SIM_STATE, OPEN289_SIM_AUDIT (file overrides). OPEN289_NO_BLE=1 makes the real Bluetooth backend refuse to start (tests set it, so a test can never reach hardware).

CLI reference

Global options go before the command:

Option Meaning
--address, -a ADDRESS adapter BLE address (default: the first IR3000 FC found)
--sim use the virtual IR3000 + Fluke 289 (no hardware)
--text human-readable output instead of JSON
--verbose, -v log adapter control-point traffic to stderr
--max-wall MAX_WALL time limit in seconds for one-shot commands (default 120; memory 600). live, log, recording and read --every have no wall limit
--version print the version

open289 <command> --help lists the options of one command.

Command Purpose Options (defaults)
scan find IR3000 FC adapters --seconds SECONDS (10)
info adapter and meter identity, battery, memory level, clock, saved counts
read full display state now (function, range, modes, all readings) --every EVERY (one snapshot per interval, e.g. 1s), -n, --count COUNT (with --every: stop after N snapshots), --for FOR_ (with --every: stop after this long, e.g. 5m)
live real-time readings from BLE notifications (~4/s), one JSON line each -n, --count COUNT, --for FOR_ (e.g. 30s), --raw (include the raw bytes), --timeout TIMEOUT (10 s without a reading is an error)
log log readings to a CSV / JSONL file --out, -o OUT (data/log_<time>.csv), --format {csv,jsonl}, --duration, -d DURATION (forever), --every EVERY (0), --snapshot-every SNAPSHOT_EVERY (0), --status-every STATUS_EVERY (10s), --reconnect-for RECONNECT_FOR (5m)
memory saved measurements, MIN MAX, peak sessions and recordings --detail (with --text: list every reading)
recording download trend recording INDEX to CSV / JSONL index, --out, -o OUT (data/recording<INDEX>_<time>.csv), --format {csv,jsonl}, --start START (0), --limit LIMIT (end index), --no-batch
press simulate front-panel keys KEY [KEY ...], --held, --settle SETTLE (1.0 s), --allow-off
raw raw meter commands (ID, IM and Q* unless --unsafe) COMMAND [COMMAND ...], --unsafe
selftest read-only health check of adapter, link and meter
stop end a running live / read --every / log cleanly

Durations (--every, --for, --duration, --snapshot-every, --status-every, --reconnect-for) take ms, s, m (or min), h, d; a bare number is seconds: 500ms, 30, 90s, 5m, 1.5h, 2d. forever (also inf, until-stop) means no limit (for --duration).

stop writes a request file that a running live, read --every or log (in any process) sees within about a second; it does not need the adapter. recording and the one-shot commands do not react to it.

Output format

Every command prints JSON on stdout, one object per line, compact. One-shot commands print one object. live, read --every, log and recording print event lines and end with an "event": "done" summary. Event lines carry "ok": true, "cmd": ..., "event": ... (started, connected, status, progress, link_lost, reconnected, done); live readings are {"type":"reading", ...}. The first field of a result is ok. Warnings and --verbose traffic go to stderr. Values are in SI base units; display is the text from the meter. Times: t, pc_time, started, ended are Unix seconds on the PC; time, start, end, info.meter.clock, and the numeric ts / clock_epoch, are the meter's local clock (no timezone).

Errors are one object and a non-zero exit code:

{"ok":false,"cmd":"read","error":{"code":"METER_BUSY","message":"'QDDB' -> EXECUTION_ERROR","hint":"the meter is on a menu, result page or setup screen: close it on the meter (or with open289 press), then retry"}}

--text prints a human-readable form of the same data: one line per reading for live (HH:MM:SS.mmm value unit and the secondary after |), a short table for read, memory, press and selftest, one line per record for recording, and a short summary for log. Errors in --text are error CODE: message plus a hint: line. Scripts and agents should use the JSON.

Errors and exit codes

Every error has a code, a message and (mostly) a hint; the exit code follows the code.

Code Exit Meaning and typical fix
INVALID_ARG 2 an argument cannot be parsed or is out of range (duration, key, index): fix the command
NOT_FOUND 3 no IR3000 FC is advertising: wake the adapter (button until the LED flashes green), come closer, close Fluke Connect on phones
BUSY 4 another open289 process holds the adapter (wait, or open289 stop); in the MCP server also a running background log
METER_BUSY 5 the meter answers EXECUTION_ERROR: a menu, result page, save page or setup screen is open; close it on the meter or with press F4
DETACHED 5 the adapter reports no powered meter (per the Fluke app it then reports FLUKE PC3000; not observed on hardware): seat it on the IR port, switch the meter on
NO_DATA 5 the meter has no data for the request, e.g. no saved item with that index: check memory
TIMEOUT 6 a one-shot command passed --max-wall, or no reading for --timeout seconds: move closer, check the meter is on, raise --max-wall
BLE_ERROR 6 scan or connect failed: is Bluetooth switched on? (also: OPEN289_NO_BLE is set)
LINK_LOST 6 the Bluetooth link dropped (a log reconnects by itself within --reconnect-for)
PROTOCOL_ERROR 6 an unexpected or undecodable answer; a recording download names the record to resume from
METER_ERROR 6 the meter did not accept the command (syntax error) or did not answer
WRITE_FAILED 6 reserved (not emitted yet; a failed Bluetooth write currently surfaces as INTERNAL)
UNSAFE_OPERATION 7 refused by the safety rules: a non-query raw without --unsafe, ONOFF without --allow-off, a send over 87 bytes
STOPPED 0 reserved; a stopped run ends normally with stopped_by: "stop"
INTERNAL 1 unexpected exception (the result carries the last trace lines); anything else maps to exit 1

Exit 0 is success; Ctrl+C gives 130. A failed selftest (a check failed, it is not an error object) and a memory where some items could not be read exit 1 and 6 respectively. The MCP server uses the same codes and adds TIMEOUT for tool calls past max_wall, CANCELLED and SELFTEST_FAILED.

MCP server

open289-mcp (or python -m open289.mcp_server) serves the meter over the Model Context Protocol (stdio, standard library only; installed with pip install open289):

claude mcp add open289 -- open289-mcp

If pip's Scripts folder is not on PATH, give the full path to open289-mcp. Other clients: {"mcpServers": {"open289": {"command": "open289-mcp"}}}. Each tool call runs one CLI command in a child process, so the adapter is connected and released per call, a hang never blocks the server, and every rule of the CLI applies (safety filter, adapter lock, error codes). The server's own instructions tell the agent to start with status.

11 tools:

Tool Runs Key arguments (defaults)
status info max_wall. Adapter address / firmware / battery, meter model / firmware / serial / battery / memory level / clock, saved counts. meter.attached false = not on a powered meter
read read max_wall. One display snapshot: function, range, modes, primary and every other reading
live live count (20, 1..400), seconds (0.1..120; with count: whichever comes first), max_wall. A summary instead of the lines: count, rate, first / last time, min / max / mean of the primary per unit, functions and units seen, the last 5 readings
memory memory max_wall (600). Counts and every saved item
download_recording recording index (required), out, format, start (0), limit, max_wall (900). File, number of records, seconds per record, retries, the recording header. A failed result carries resume with the start and out to continue with
press press keys (required, 1..20 of F1 F2 F3 F4 UP DOWN LEFT RIGHT HOLD MINMAX RANGE INFO BACKLIGHT ONOFF), held (false), allow_off (false), max_wall. The display state after each key
query raw commands (required, 1..10, e.g. ["ID","QM"]), max_wall. Read-only: ID, IM, Q*; anything else is refused with UNSAFE_OPERATION before it reaches the meter (there is no --unsafe in MCP)
selftest selftest max_wall. The checks S01-S09; a failure returns SELFTEST_FAILED with the failed checks
log_start log in the background duration (10m; forever = until log_stop; must be > 0), every (0), snapshot_every (0), out, format, reconnect_for (5m). Returns a job id and the file
log_status none (reads the job's output) job (the latest). State (running / done / failed), records so far, the last status line, the file, and when ended its summary
log_stop stop job (the latest). Stops cleanly and returns the summary: file, records, duration, min / max / mean per unit, link gaps. Without job it also stops a log or live started elsewhere (a terminal, another client); already_ended: true = it had ended before

status, read, live, memory, query, selftest, log_status are marked read-only. All durations are strings ("30s"), all numbers are SI.

Background logging. log_start returns at once (after the first reading, at most 30 s) with a job id; the log keeps running when the client disconnects, and its output and metadata are kept in state/mcp_jobs/<job>/, so a restarted server, or another client's server, still sees it. Follow it with log_status, end it with log_stop (or it ends after duration, or when the link stays lost longer than reconnect_for). The file holds the data; the tool results are summaries and never hundreds of readings.

BUSY rules. The adapter accepts one connection, so tools run one at a time, and while a background log runs, every other meter tool (status, read, live, memory, press, query, ...) returns BUSY until log_stop (log_status and log_stop still work). BUSY also means another open289 (a terminal running live/log, another client) holds the adapter. Fluke Connect on a phone shows up as NOT_FOUND (the adapter stops advertising). METER_BUSY is different: a menu or result page is open on the meter; close it (press F4) and retry. Safety for agents: press changes the meter (HOLD, MIN MAX, range); ONOFF needs allow_off: true and should be used only when the user asked for it: it switches the meter off for good, and the adapter goes to sleep with it.

CLI commands with no MCP tool: scan (the server finds the adapter itself), stop (used internally by log_stop), and raw --unsafe (deliberately not exposed). An agent that has a shell can use the CLI for all of them.

Data folders

open289 keeps its files in one folder with two subfolders:

  • state/: the adapter lock (open289.lock), the stop request (stop.request), MCP background jobs (mcp_jobs/) and the virtual meter's state (sim_state.json, sim_writes.jsonl).
  • data/: the default place of logs (log_<time>.csv) and recording downloads (recording<INDEX>_<time>.csv). --out writes anywhere else; relative paths are relative to the current folder.

Where the folder is: OPEN289_HOME if set; else, in a source checkout (a pyproject.toml next to the package), the checkout itself; else, for an installed package, %LOCALAPPDATA%\open289 on Windows and ~/.local/share/open289 elsewhere, so nothing is written into site-packages. The MCP server's child processes use the same folder.

Development

Layout:

  • open289/ package: cli.py (commands, output), transport.py (backend selection real or virtual, discovery, connection), adapter.py (IR3000 service client, pass-through, the opcode allowlist, query filter and the 87-byte limit), meter.py (the meter's remote commands and the binary record decoders), reading.py (live notification decoder), records.py (JSON-ready dictionaries), logger.py (long-running log with reconnect), lock.py (single client lock), errors.py, paths.py, sim.py (virtual meter) and mcp_server.py (MCP server).
  • tests/ offline tests against the virtual meter: decoders (against real captures), the transport and its safety rules, the CLI, the MCP server (JSON-RPC, jobs, BUSY rules) and the simulator itself.
  • docs/PROTOCOL.md (and PROTOCOL.zh-CN.md): the protocol, with evidence levels.
  • tools/ diagnostic scripts for the hardware.
  • .github/workflows/: ci.yml and release.yml.
python -m pip install -e ".[test]"
OPEN289_SIM=1 OPEN289_NO_BLE=1 python -m pytest tests -m "not hw" -q

The tests set the simulator environment themselves (tests/conftest.py: a private OPEN289_HOME per test, OPEN289_SIM=1, OPEN289_NO_BLE=1, a fast simulator), so a test can never reach Bluetooth. Tests marked hw need a real adapter and meter and are skipped by default (pytest -m hw).

CI (ci.yml, on pushes to main, pull requests and by hand): the tests on Windows and Linux with Python 3.10, 3.12 and 3.14, all against the virtual meter; a check that tracked files carry no local paths; a package build with twine check --strict and a smoke test of the built wheel (open289 info, open289 live -n 5 on the virtual meter).

Release (release.yml): push a tag vX.Y.Z equal to open289.__version__ (v0.1.1 for 0.1.1). The workflow runs the CI checks, refuses a tag that does not match the version, builds, creates a GitHub release with the files, and publishes to PyPI by trusted publishing (no token). One-time setup on pypi.org: add a (pending) trusted publisher for the project open289 with owner FrostFlops, repository open289, workflow release.yml and environment pypi, and create the environment pypi in the repository's settings.

Protocol

How the adapter and the meter talk (GATT table, the CNX ID write that arms the control point, the pass-through framing, buffer limits, the 16-byte live packet layout, the binary saved-data records, key behaviour), with every statement tagged VERIFIED (seen on the hardware) or DERIVED (not exercised): docs/PROTOCOL.md (中文). Observed live encodings per function: section 4.8 of docs/PROTOCOL.md.

What is NOT here

Fluke's application (Fluke Connect), anything decompiled from it, and the private reverse-engineering workspace are kept out of this repository. The firmware images and updater of the adapter are not used.

Safety

  • The tool never writes the adapter's firmware-update characteristics (the MSP updater and the radio-image updater; writing the wrong value there erases flash) and never sends the adapter opcodes 0x90 to 0x94 (the 1550 / 1555 insulation-test commands, which switch a high voltage).
  • The adapter's out-buffer is capped at 87 bytes per send: larger sends got no reply or hung the adapter until it was power-cycled (373 bytes and more), so open289 refuses them.
  • raw is query-only (ID, IM, Q*) unless --unsafe. MP can change settings, CSD erases saved data, RI resets the meter, OFF switches it off.
  • press ONOFF needs --allow-off, and the meter asks for confirmation (F1). OFF and ONOFF cannot be undone remotely: neither the meter nor the adapter can be switched on again from the PC.
  • press otherwise changes meter state (HOLD, MIN MAX, range, saving a measurement on the save page): it is not read-only, and an agent should only use it on request.
  • The adapter accepts one Bluetooth connection: close Fluke Connect on phones. open289 holds a lock so that only one open289 process connects at a time.
  • This is a measurement reader, not a safety device. Respect the meter's category rating and its own safety rules; open289 does not make a measurement safer.

License

GPL-3.0-only (see LICENSE). Fluke, Fluke Connect and IR3000 are trademarks of Fluke Corporation; this project is not affiliated with Fluke Corporation.

Metadata

Release files for open289 0.1.1

For a detailed explanation of source distributions (sdists) and built distributions (wheels), please see the package formats documentation.

Source distribution (sdist)

Source distribution for open289 0.1.1
File Size Uploaded Provenance Actions
open289-0.1.1.tar.gz 171.1 kB With provenance Details for open289-0.1.1.tar.gz

Built distribution (wheel)

Table of built distributions (wheels) for open289 0.1.1
File Interpreter ABI Platform Actions
open289-0.1.1-py3-none-any.whl Python 3 none any Details for open289-0.1.1-py3-none-any.whl

Total release size: 287.2 kB

Release files / open289-0.1.1.tar.gz

Download URL open289-0.1.1.tar.gz
Size 171.1 kB
Tags Source
SHA-256 checksum
How to use checksums
5e29876e4de1d3d5b78dd8a6d8e780f9cbc9ac8453467a829d076e875562eeb7
BLAKE2b-256 checksum
How to use checksums
5dd141c0f372a52176d5d7561bf2cea33e1d3b565d96ae25cf50589185750174
Upload date
Uploaded using Trusted Publishing?
What is trusted publishing?
Yes
Uploaded via twine/7.0.0 CPython/3.13.16

Provenance

Provenance describes where a file came from. On PyPI, provenance is shared via attestations, which provide a verifiable record of the build or publishing details. View details, limitations and caveats.

PyPI Publish Attestation

PyPI verified that this artifact, at this checksum, originated from the publisher listed below.

Signed by GitHub Actions, verified by PyPI on Nov 7, 2026.

Transparency log

Release files / open289-0.1.1-py3-none-any.whl

Download URL open289-0.1.1-py3-none-any.whl
Size 116.1 kB
Tags Python 3
SHA-256 checksum
How to use checksums
0e990c016c183c171a701820807d72a6e5ad912136ea962986b738a551149fdc
BLAKE2b-256 checksum
How to use checksums
50db2004b32ce3003a9ac074c9349172059980d4d7d76e50818ee158258c654d
Upload date
Uploaded using Trusted Publishing?
What is trusted publishing?
Yes
Uploaded via twine/7.0.0 CPython/3.13.16

Provenance

Provenance describes where a file came from. On PyPI, provenance is shared via attestations, which provide a verifiable record of the build or publishing details. View details, limitations and caveats.

PyPI Publish Attestation

PyPI verified that this artifact, at this checksum, originated from the publisher listed below.

Signed by GitHub Actions, verified by PyPI on Nov 7, 2026.

Transparency log

Security for open289 0.1.1

Provenance

Provenance describes where a file came from. On PyPI, provenance is shared via attestations, which provide a verifiable record of the build or publishing details. View details, limitations and caveats.

Full provenance

Both files in this release have been uploaded with provenance metadata.

Select from 2 files

Attestation for open289-0.1.1.tar.gz

PyPI Publish Attestation

PyPI verified that this artifact, at this checksum, originated from the publisher listed below.

Signed by GitHub Actions, verified by PyPI on Nov 7, 2026.

Transparency log

A note on package security

PyPI is a repository for distributing packages, not a service for auditing them. While third-party security researchers and our staff do their best to detect and remove malicious packages, PyPI cannot guarantee the quality, security or suitability of any given package.

To learn more about protecting yourself from malicious actors, visit our package security documentation.

Release history Release notifications | RSS feed

This release

0.1.1 This release

2 release files Full provenance

Anthropic, PBC Visionary sponsor Bloomberg Visionary sponsor Hudson River Trading Visionary sponsor Meta Visionary sponsor NVIDIA Visionary sponsor Microsoft Sustainability sponsor Depot Continuous Integration AWS Cloud computing and Security Sponsor Datadog Monitoring Fastly CDN Google Download Analytics Sentry Error logging StatusPage Status page