openm454
English | 简体中文
An open driver, command-line interface and MCP server for the DreamSourceLab DSCope M454 USB mixed-signal oscilloscope (4 analog channels up to 400 MSa/s, 8 digital channels, built-in 0-3 V signal generator), written so that an AI agent (Claude Code, or any MCP client) can capture and analyze real signals by itself, without the vendor GUI.
The M454 speaks a newer ("v3") USB protocol than the devices supported by the open-source DSView. The protocol was
recovered from DSWave 1.2.1 by static analysis, by driving DSWave's own hardware layer through injected commands and
by USB tracing; see docs/PROTOCOL.md.
Use
Requirements: Windows, Python 3.10+, the scope bound to WinUSB (it is by default), a USB 3 port and cable (captures
refuse a USB 2 link with UNSUPPORTED), and DSWave 1.2.1, needed once to
extract the FPGA bitstream (it is never redistributed). Close DSWave before using the driver: the scope accepts one
USB client.
pip install openm454 # or, from a clone: pip install -e .
The openm454 data folder is the clone itself, or for an installed copy %LOCALAPPDATA%\openm454
(~/.local/share/openm454 elsewhere; OPENM454_HOME overrides it). It holds the cached bitstream (firmware/),
state (state/) and captures (captures/).
Where the FPGA bitstream comes from
DSWave 1.2.1 (DSWave.exe) is needed once. openm454 only reads its files, looks in these places in order (each may
be the folder holding DSWave.exe or the exe itself) and searches no disk beyond them:
- the environment variable
OPENM454_DSWAVE(DSCOPE_DSWAVEis accepted too); - the path saved in
state/dswave.jsonof the data folder (written byopenm454 setup --dswave, below); DSWave/in the data folder (a copy of the DSWave folder);- the Windows uninstall entries of DSWave in the registry: a normal DSWave installation is found with no setup at all;
%ProgramFiles%,%ProgramFiles(x86)%and%LOCALAPPDATA%\Programs, each inDSWave\orDreamSourceLab\DSWave\.
The bitstream is extracted the first time the scope's FPGA has to be configured (the first command after the scope is
powered up) or by setup --dswave, checked by SHA-256 and cached in firmware/ of the data folder; DSWave can then be
closed, moved or uninstalled. setup without --dswave only reports (cached, dswave, source, ready). If nothing is found, the error is FIRMWARE_SOURCE (exit 6)
with what was searched and what to do. To see the state, or to tell openm454 where DSWave is once (no scope needed):
python -m openm454 setup # cached? where is DSWave found, and how
python -m openm454 setup --dswave <folder> # check DSWave 1.2.1 there, cache the bitstream, save the path
The saved path is read, with no restart, by every openm454 process that uses the same data folder (the MCP server
included: if it was registered with -e OPENM454_HOME=<folder>, run setup with the same OPENM454_HOME).
python -m openm454 info
python -m openm454 capture -c A1,D0 --vdiv 1V --offset 1.5V --timebase 200us --trig A1:rise:1.4V
python -m openm454 watch -c A1 --every 0.5s --count 10 --fields A1.freq_hz,A1.vpp_v
python -m openm454 awg set --wave sine --freq 10k --low 0V --high 2V
python -m openm454 analyze captures/<file>.npz --window=-1ms:1ms
python -m openm454 capture -c D0,D1 --rate 20M --duration 2ms --trig D0:pulse>:200us --decode "i2c:scl=D0,sda=D1"
python -m openm454 selftest
Recording modes and measured speeds
| Command | How it records | Length | Rates |
|---|---|---|---|
capture |
into the 512 MiB sample memory, read out afterwards (250-265 MB/s) | one record: 400M samples per channel / 512 MiB | up to 400 MSa/s (1 channel) |
stream |
every record is sent while it is acquired, at any data rate (--record-first: above 102.4 MB/s acquire it first and read it out afterwards, as DSWave does); longer than one record or --duration forever: record after record, each refreshing the previous one |
unlimited; between records ~0.2 s while the readout keeps up, plus the rest of the record above it (0.7 s at 250 MB/s; 2.4 s with --record-first); the last record is kept (--save-each: all) |
as capture: analog and digital at their own rates |
log |
loop mode: one endless gap-free stream, kept in RAM (default; written once at the end; a RAM ring keeps the newest part of longer runs) or, with --storage disk, dumped in chunk files while running |
until openm454 stop or the duration; RAM budget / disk allowance |
one rate for all channels, 100 MHz / N for any integer N (1 kSa/s..100 MSa/s, e.g. 33.333 MSa/s); interleaved 200 MSa/s (<= 2 analog, <= 4 digital channels) and 400 MSa/s (<= 1 analog, <= 2 digital) |
Measured on the M454 over USB 3 with every input on a 1 kHz square, every period checked for gaps (2026-11-01/02):
- The data rate is rate x (analog channels + digital channels / 8) bytes/s.
stream: real time at DSWave's 100 MB/s ceiling and at 200 MB/s. At 400 MB/s the readout ran at 270-290 MB/s and the rest waited in the sample memory; the record arrived complete. 12 channels at 50 MSa/s (250 MB/s, ~2 s records): 0.74 / 0.66 s between records when sent while acquired (the default), 2.35 / 2.45 s with--record-first; records complete and continuous.log: 150 MB/s (2 analog + 8 digital at 50 MSa/s) for 3 min (27 GB) and for 30 s to disk with a constant host lag; 200 MB/s (2 analog at 100 MSa/s) for 20 s without backlog; 5 MB/s for 10 min. At 250 MB/s (4 analog + 8 digital at 50 MSa/s) the readout reached ~200 MB/s and the backlog grew ~55 MB/s. A 300 MB/s burst of 3 s (2 analog + 8 digital at 100 MSa/s) read out at 266 MB/s and stayed complete.logabove 100 MSa/s interleaves the converters exactly ascapturedoes (DSWave never does this in loop mode). Verified on the scope: 1 analog channel at 200 and 400 MSa/s, 2 analog at 200, 2 digital at 400, 4 digital at 200 and mixed sets. 2 digital channels at 400 MSa/s or 4 at 200 MSa/s are 100 MB/s and run without a time limit. 1 analog channel at 200 MSa/s (200 MB/s) ran 10 s without backlog.- The sample memory is a 512 MiB ring in loop mode: the host may fall up to ~512 MiB behind without losing data (host
stalls of 10 s at 50 MB/s were absorbed). Beyond that the scope overwrites unread data without any flag and keeps
streaming: from then on the data are garbage (one phase step on analog, the digital channels misaligned) until
the run is restarted. Measured with the driver's backlog estimate, the break came at 508-514 MiB (three runs), so
logstops at 480 MiB. log --plan(no device needed) prints the verdict for a rate, the gap-free limit and recommendation (next section), how long the run can last (max_duration_s) and where the data goes: RAM by default (refused when it does not fit the budget unless--keep-last),--storage diskfor ~256 MB chunk files (it warns about SSD wear: 200 MB/s is 720 GB per hour),--storage nonefor statistics only.benchmeasures this PC (free RAM, decode speed, disk write speed, USB readout) and records how much RAM and disk the owner allows for recordings and the data folder (bench --ram 8GB --disk 50GB --dir <folder>,state/host.json);logtakes its defaults from there. Measured here: decode 468 MB/s, USB readout 206 MB/s (4 analog + 8 digital) / 267 MB/s (2 analog + 8 digital).
Above the readout limit: how log decides
D = the data rate; R = how fast this PC reads the scope while it acquires. R depends on the channel set (measured here: 202 MB/s for 4 analog + 8 digital at 50 MSa/s, 264 MB/s for the same at 100 MSa/s and for 2 analog + 8 digital at 100 MSa/s, 275 MB/s for 4 analog at 100 MSa/s).
- R is measured, not assumed: the first
logabove 150 MB/s with a channel set runs a probe of up to 2 s (aprobeline) and remembers R instate/host.json: the measured rate, or "kept up" (R >= D) when it reads at= 99 % of D. Later runs above 150 MB/s refresh it when the result is clear (still falling behind at a steady rate at the stop, or caught up at the end), and so do
benchand--remeasure. Until then--planassumes 190 MB/s and marks the planprovisional. - D <= R: the host keeps up (
sustained/near_limit); the run lasts as long as the storage allows. - D > R: the scope gets D - R ahead per second (
backlog_growth_mb_s), and it sends nothing after a stop, so the unread part is lost. A log is gap-free for at mostgapfree_max_s= 480 MiB x R / (D x (D - R)), the signal the host has read when the scope is 480 MiB ahead. A shorter log runs normally. - A longer log (or
forever) is compared with whatstreamwould do with the same channels and settings (stream_alternative, from stream's own planning): records of up to 400M samples per channel / 512 MiB, sent while acquired, with gaps of record x (D - R) / R + 0.18 s, at stream's rates (1-2-5 steps x interleave; it may run a domain slower, e.g. 2 analog + 8 digital at 100 MSa/s samples the digital channels at 50 MSa/s):STREAM_BETTER(exit 2): stream samples every channel at least as fast, and its records are at least as long as the log's one gap-free piece (or one record holds the whole duration). The error carries the stream command, with the same vertical settings.NEEDS_FORCE(exit 2): the log's piece is the longest gap-free recording possible but shorter than asked.--forcerecords it and then stops (stop_reasonring_full, ok, with a warning); the rest of the requested time is not recorded. Both errors carry the numbers inerror.plan.
- While running, every statistics line has
backlog_mb; while the backlog grows, alsobacklog_growth_mb_s,ring_full_in_sandgapfree_projected_s. The result hasunread_at_stop_mb(acquired, never read).
Examples here: 12 channels at 50 MSa/s (250 MB/s, R = 204): gap-free 8.88 s vs stream records of 2.03 s with 0.64 s
gaps, so NEEDS_FORCE; forced, the run stopped after 8.84 s (8.30 s with R = 202 in an earlier run), every period
present up to the last sample. 12 channels at 100 MSa/s (500 MB/s, R = 264): 1.13 s vs 1.02 s records with 1.09 s
gaps, still NEEDS_FORCE; with a readout below ~250 MB/s stream would win. 200 MB/s read at 190 MB/s: 10 MB/s
backlog growth, the guard after ~50 s, 47.8 s of gap-free signal.
The backlog is estimated as D x (time since the head) - bytes received, checked on every USB transfer, against a
baseline that follows the scope / PC clock difference: it may rise by at most 200 ppm x D per second, and only
while the host is caught up (estimate below 16 MiB), so the estimate stays exact over hours and a real backlog is
never absorbed. A guard stop while nothing arrives at all is reported as a stalled scope (TIMEOUT_DATA); a
forced log that was planned to end at the guard reports ring_full, any other guard stop HOST_TOO_SLOW.
log or stream?
Both record untriggered, but they are not the same mode of the scope. log is one endless run (loop mode);
stream is a series of records, each programmed with a fixed length that the scope stops at.
log |
stream |
|
|---|---|---|
| gaps | none inside a run | one per record (refresh) |
| gap between two runs, 12 channels at 20 MSa/s (100 MB/s) | 0.18 s | 0.19 s |
| gap between two runs, 12 channels at 50 MSa/s (250 MB/s, above the readout limit) | 0.51 s after a 1.5 s run: 0.33 s to read the run's backlog (data the scope acquires meanwhile are dropped), then 0.18 s restart | 0.66-0.74 s after a ~2 s record: the rest of the record is read (sent while acquired), then the restart; 2.33-2.45 s with --record-first (the whole record read out after it was acquired) |
| rates | one rate for every channel: 100 MHz / N for any integer N, interleaved 200 / 400 MSa/s on 1-2 channels | analog and digital at their own rates (e.g. 1 analog at 400 MSa/s with 8 digital at 100 MSa/s), from DSWave's 1-2-5 x interleave table (e.g. 80 MSa/s) |
| above the readout limit (~200-275 MB/s) | one gap-free piece of at most 480 MiB x R / (D x (D - R)) (the scope sends nothing after a stop; see above) | records of up to 512 MiB / D (400M samples per channel), each sent while acquired, with gaps |
| below ~51 kB/s (e.g. 1 channel at 1-20 kSa/s) | supported | the rate is raised |
| protocol decoders | offline on the saved file (analyze --decode) |
--decode on the kept record |
| register program | DSWave's Data Acquisition mode (interleaving: this driver's extension, verified) | DSWave's own program for this window (golden-tested) |
The restart (~0.18 s) is the same for both; the difference is what has to be read out before it. Use log for
continuous recording; stream for different analog and digital rates, decoding the record, or above the readout
limit when log says STREAM_BETTER or the whole time matters more than one long gap-free piece.
python -m openm454 bench --ram 8GB --disk 50GB --dir logs
python -m openm454 log --plan -c A0,D0-D7 --rate 1M --duration 10min
python -m openm454 log -c A0,D0-D7 --rate 1M --duration 10min # kept in RAM, one .npz at the end
python -m openm454 log -c A0,D0-D7 --rate 1M --duration forever # RAM ring; end it with: python -m openm454 stop
python -m openm454 stream -c D0-D7 --rate 100M --duration 60s --every 1s
Every command prints one JSON document (watch, stream and log: one per line, then the result), whether or
not --json is given: measurements, applied settings, file paths. Samples are never printed (unless --dump N asks
for a decimated series); captures are saved as .npz (raw codes + scaling) and .png. In the plot the analog legend is one row above the plot area (one entry per
channel, e.g. A1 1V/div Vpp 2.86V f 1kHz) with the title above it, digital lane labels sit in the right margin,
and only the trigger markers (dashed line at t = 0, dotted analog trigger level) are drawn over the waveform; decoded
frames are not drawn, they are in the JSON and <capture>_decode.json. python -m openm454 <command> --help lists every option. Python API: openm454.scope.Scope, openm454.capture.load_capture, Capture.decode(), openm454.plot.render(cap, window=None, title=None, summary=None, channels=None) (returns the figure; the caller closes it), openm454.firmware.status() / save_dswave().
Protocol decoders (--decode SPEC, repeatable, on capture, watch, stream and analyze):
uart:rx=D4,baud=115200|auto,bits=8,parity=none,stop=1,invert=0, spi:clk=..,mosi=..,miso=..,cs=..,mode=0..3,
i2c:scl=..,sda=.., can:rx=..,bitrate=500k|auto,invert=0. They report every frame with trigger-relative times and
its errors (parity/framing, partial words, NACKs, CRC/stuff/form errors), and remove the comparator chatter of slow
edges. Output format: docs/schemas/decode.json. Decoding needs about 10 samples per bit or clock (I2C: 9 per SCL
period, no errors; 5-6 per period still decode with well under 1 % errors). They have been checked against real buses (SPI up to 18 MHz, RS232, CAN up to 1 Mbit/s, I2C up to 400 kHz)
generated by a microcontroller, including deep captures and decoded streams.
CLI reference
pip install openm454 installs the openm454 command (python -m openm454 ... is the same) and the openm454-mcp
server. Global options, before the subcommand: --max-wall T (watchdog: stop and release the device after T or the
budget computed from the command, whichever is longer; it never shortens the computed budget), --verbose (wire
log on stderr), --json (accepted, output is always JSON).
python -m openm454 <command> --help lists every option.
| Command | Purpose | Main options |
|---|---|---|
info |
device, firmware, calibration, limits | |
capture |
one triggered (or untriggered) record: measurements, .png, .npz |
see below |
watch |
repeated captures, one JSON line each (monitoring) | the capture options, --every 0.5s (start to start), --count N (default 20 when neither --count nor --for is given), --for T, --fields A1.freq_hz,..., --stop-when EXPR, --save-last (see watch) |
stream |
untriggered recording sent while it is acquired, live statistics | the capture options, --every 1s, --save-each, --record-first |
log |
gap-free endless recording (loop mode), one rate for all channels | -c, --rate 1M, --duration 10s|forever, --every 1s, --storage ram|disk|none (--no-save = none), --keep-last, --chunk, --ram, --plan, --force, --remeasure, --out, --tag, --awg, --hook FILE.py (live processing, see Live processing hook), the vertical options |
bench |
this PC's recording limits (free RAM, decode and disk speed, USB readout); stores the RAM / disk allowance | --ram 8GB, --disk 50GB, --dir, --disk-test-mb, --no-device |
awg set |
continuous waveform on the MCX output (0..3 V) | --wave sine|square|triangle|noise, --freq 1k, --low/--high or --vpp/--dc, --duty 50%, --phase |
awg sweep |
frequency sweep, repeating | --start, --end (required), --time 100ms, --log, --wave, --high, --low, --duty |
awg burst |
N periods of the carrier every period | --cycles, --period 10ms, --freq, --wave, --high, --low, --duty |
awg mod |
AM / FM / FSK of the carrier | --kind am|fm|fsk, --mod-wave, --mod-freq, --depth, --deviation, --hop, --freq, --wave, --high, --low, --duty |
awg dc |
steady DC level (0..3 V, 6.4 mV steps) | --level (required) |
awg off, awg status |
park the output at 0 V / show the last setting | |
plot |
render a saved .npz to PNG (no device) |
file, --window, --channels, --out |
analyze |
measure a saved .npz (no device) |
file, --window, --channels, --edges N, --dump N, --measure, --decode SPEC |
selftest |
verify the scope (Self-test) | --quick, --no-awg, --fixture file|standard |
setup |
where the FPGA bitstream comes from; --dswave <folder> records DSWave's folder once (no USB) |
--dswave |
release |
make sure the device is idle and released | |
recover |
reset and reload the FPGA of a hung scope, like a power-up (a scope that does not answer register reads still needs a replug) | |
stop |
ask a running watch / stream / log (e.g. in the background) to stop cleanly |
|
debug |
maintenance: regs-read, regs-write (same safety policy), plan (print the register program for capture options without touching the device) |
capture, watch and stream share these options:
- channels / vertical:
-c A0-A3,D0-D7(alsoanalog,digital),--vdiv 1V,--offset 0V,--coupling dc|ac,--bw20 A1,A2,--invert A2,--probe 10x|A2=10x,--threshold 1.4V. Per-channel values are writtenA0=500mV,A2=2V. - timing:
--timebase(window = 10 divisions),--duration,--samples,--rate. - trigger:
--trig SRC:EDGE[:LEVEL](e.g.A1:rise:1.4V,D0:fall,D0:pulse>:200us),--trig-mode auto|normal|force(defaultauto),--trig-pos 50%,--auto-after 0.3s,--timeout 5s(normal-mode wait),--holdoff,--sensitivity 0..15(default 5),--trig-retries N(default 4; digital edge triggers, see Digital edge triggers).streamaccepts the trigger options and ignores them: it always records untriggered. - output:
--measure basic|full|none(see Measurements),--edges N,--dump N,--window t0:t1,--no-plot,--no-save,--out,--tag,--awg 'sine,10k,0V,2V[,duty%]'(set the generator first),--decode SPEC(repeatable),--decode-limit(default 50 frames).
log takes the vertical options too; its timing is --rate and --duration (see the recording sections above).
A value that starts with - needs the = form on Python 3.10-3.12 (--window=-1ms:1ms, --offset=-1.5V; per
channel A1=-1.5V is fine as it is); the MCP server does this itself. --trig without a level on an analog channel
uses that channel's --offset as the level. --rate snaps down to a rate the scope has (at most 400M on 1 analog
channel, 200M on 2, 100M on 3-4; digital only: 400M on up to 2 channels, 200M on up to 4, 100M on 5-8).
capture keeps only the newest 200 captures with its --tag (default cap) in its output folder (captures/ by
default, or --out): older files named the way openm454 names them (<yyyymmdd-hhmmss-mmm>_<tag>.npz, .png,
_decode.json) with that tag are deleted. Recordings with other tags (log, stream, watch_last) and any other file
in the folder are never touched.
Exit codes: 0 ok, 1 other (a failed self-test, an internal error, interrupted), 2 bad arguments, 3 scope not found, 4 busy, 5 no trigger, 6 device / protocol / watchdog failure, 7 refused by the safety rules (see Errors).
watch
watch repeats capture and prints one JSON line per capture.
--every T(default 0.5s) is the time from the start of one capture to the start of the next. If a capture takes longer, the next one starts at once, with no catch-up burst.t_sof a line is when that capture started, in seconds since the watch began (--every 0.5sgives 0, 0.5, 1.0, ...). The final line haselapsed_s.--count Nand--for Tend the watch; with both, whichever comes first. No capture starts at or after--for:--every 0.5s --for 2sgives 4 captures. With neither, 20 captures.openm454 stopends the watch after the current capture, or within 0.25 s while it waits between captures.--fields CH.KEY,...:KEYis any measurement name; only the named channels are measured. The default fields arevpp_v,vmean_v,freq_hz,duty_pctandstateof every channel.--stop-when EXPR: a Python expression with one dict per channel, e.g."A1['vpp_v'] < 1"; an expression that raises counts as false.--fieldsor--stop-whentogether with--measure noneis refused (exit 2).--windowand--edgesapply;--dump,--no-plot,--no-saveand--taghave no effect.--outis only used by--save-last, which writes<out>/<date>-<time>_watch_last.npz.- A line adds
trigger_checkwhen it is notok/unchecked, andtrigger_misfireswhen above 0.
Measurements (--measure)
--measure basic|full|none applies to capture, watch, the final result of stream and analyze.
basic(default). Analog channels:samples,vmin_v,vmax_v,vpp_v,vmean_v,vrms_v,vac_rms_v,vtop_v,vbase_v,vamp_v,freq_hz,period_s,duty_pct,edges_rise,edges_fall,periods_in_window,pwidth_s,nwidth_s,rise_s,fall_s,overshoot_pct,undershoot_pct,clip_lo,clip_hi, and when they occurfreq_method,period_irregular,noise_like. Digital channels:samples,state,high_frac,edges,freq_hz,period_s,duty_pct, and when they occurglitch_edgesandrunts. On analog channels, frequency and duty come from falling edges when the window holds at least two falls but fewer than two rises (digital channels need two rising edges).full:basicplus, per channel, statistics over every whole cycle (rising edge to rising edge) in the window:cycles,period_min_s,period_max_s,period_std_s,freq_min_hz,freq_max_hz,pwidth_min_s,pwidth_max_s,nwidth_min_s,nwidth_max_s,duty_min_pct,duty_max_pct(digital channels also get the mean widthspwidth_s,nwidth_s), andcycles_capped: truewhen the 100k-edge limit was reached (the statistics then cover the first 100k edges). Not given for channels with fewer than 2 rising edges or for noise. Analog channels also getrange_lo_vandrange_hi_v, the voltages the screen can show at the applied V/div and offset; readings at the ends are clipped. Edge lists (edges_rise_s,edges_fall_s) appear only with--edges N. Deep records take a second measuring pass.none: noanalog/digitalblocks (hints are still given). The MCPdetailargument (summary|full) is a different thing: it chooses a compact summary or the CLI's whole document.
Digital edge triggers
The digital inputs have no hysteresis. A slow edge (for example the ~8 us rising edge of the calibration square)
chatters at the threshold for tens of ns, so a falling-edge trigger can fire inside the rising edge, and the reverse,
while the hardware still reports a valid trigger. After every capture or watch triggered on a digital channel with
a rise or fall edge (normal or auto mode), openm454 therefore checks the settled level of the trigger line just
before and after t = 0 (a level is settled when it lasts at least 200 ns and at least 3 samples, whichever is longer;
shorter runs are chatter). It only calls a capture a misfire when the unsettled stretch between the two settled levels
is as short as chatter (at most 200 ns; measured chatter lasts up to about 120 ns); a longer stretch is real data and
gives inconclusive. So fast data right at the trigger (faster than about 10 Mbit/s, e.g. a short 1-0 pair of bits
between a long low and a long high) can be taken for chatter and re-armed, and an edge so slow that its chatter runs
last 200 ns or more passes as ok.
A capture that fired on the opposite edge is discarded and the scope re-armed, up to --trig-retries N times (default
4; 0 = only check; MCP trig_retries). Re-arms share the original trigger wait (--timeout in normal mode,
--auto-after in auto mode, plus --holdoff); only time spent waiting for a trigger counts. Analog triggers, any
and pulse triggers, --trig-mode force and the untriggered fallback of auto mode are not checked.
The capture's acq block reports it:
trigger_check:ok(the first capture shows the requested edge);misfire_retried(misfires were discarded, the returned capture is correct);mismatch(the returned capture is itself a misfire: t = 0 is NOT the requested edge, with a warning and a hint);inconclusive(the edge could not be confirmed, e.g. no pre-trigger samples at--trig-pos 0, or a glitch or clock faster than 200 ns; never re-armed);unchecked(not a digital rise/fall trigger, or untriggered). MCP summaries leave outunchecked.trigger_misfires: the number of misfired captures discarded and re-armed (left out of MCP summaries when 0).attemptslists every run:triggered,forced(untriggered),misfire(fired on the wrong edge),no-trigger.acq.triggeredkeeps its meaning: the hardware trigger flag of the returned capture.- Warnings and hints:
mismatchgives a warning and a hint;misfire_retriedalso adds a hint with the number of discarded misfires (e.g. "D5 falling-edge trigger: N misfire(s) inside a rising edge were discarded and re-armed").
Measured on the M454 (1 kHz calibration square, normal mode): before the check, a D5 falling trigger fired inside
the rising edge in 13 of 40 captures at 100 MSa/s, D0 falling in 5 of 40, D5 rising never. With the check, 60 captures
each: no wrong edge returned at 50, 80 or 100 MSa/s (D5 alone, D5 with A1, all 8 digital channels; 5-25 misfires
re-armed per set). At 400 MSa/s on D5 alone about half of all tries misfire, and 2 of 60 captures misfired on all
5 tries: both were returned as mismatch with the warning, not as a success.
For a reliable digital trigger use the edge that is fast on the signal (a falling trigger misfires on a slow rise),
or trigger on an analog channel, whose trigger has a noise margin.
Vertical limits
-
--offsetis the voltage at the centre of the screen. The allowed range is about -4.32 .. +4.36 divisions (-127 .. +128 offset codes), where a division is the hardware V/div (the requested value rounded up to a 1-2-5 step, 1 mV .. 5 V at the probe input). At 1x:V/div allowed offset V/div allowed offset 1 mV -4.323 .. +4.357 mV 100 mV -0.4323 .. +0.4357 V 2 mV -8.647 .. +8.715 mV 200 mV -0.8647 .. +0.8715 V 5 mV -21.62 .. +21.79 mV 500 mV -2.162 .. +2.179 V 10 mV -43.23 .. +43.57 mV 1 V -4.323 .. +4.357 V 20 mV -86.47 .. +87.15 mV 2 V -8.647 .. +8.715 V 50 mV -216.2 .. +217.9 mV 5 V -21.62 .. +21.79 V Every number scales with the probe factor (at 10x: 10 mV .. 50 V/div at the tip, limits x10).
-
A larger offset is clamped with a warning, e.g.
A1: offset 2.5 V out of range at 0.5 V/div (allowed -2.162 .. 2.179 V, about +-4.3 div); clamped to 2.179 V; the value used is inapplied.offset_v.watchlines show no warnings. -
The trigger level must be on the screen (codes 10..245 around the offset), otherwise
TRIGGER_OUT_OF_RANGE(normal and auto mode; with--trig-mode forcean off-screen level is clamped without an error). Without a level, an analog trigger uses--offset. -
Clipping gives a hint (
A1 clips at the top: increase --vdiv or lower --offset); the front end needs ~0.2-0.3 s to recover after an overdrive.
Self-test
python -m openm454 selftest checks the scope itself. The M454 has a 1 kHz calibration output (a ~3 V square wave,
the probe-compensation signal) and the signal generator output, but no internal loopback: the inputs are not
connected to either inside the scope. The standard fixture is therefore the scope's own two outputs cabled to its
inputs, with no other instrument. The checks fall in two groups.
Wiring-free checks run on any wiring. They only read the inputs and never touch the generator (about 4-6 s):
- V01 device info
- V12 a forced capture of all 12 channels (A0-A3, D0-D7) at 100 MSa/s
- V13 the top rates and fold modes: 1 analog channel at 400 MSa/s, 2 analog at 200 MSa/s, 2 digital at 400 MSa/s, 4 digital at 200 MSa/s
- V11 capture latency
Signal checks (V00, V02-V10: calibration square, V/div, sample-rate consistency, digital channels and threshold, trigger position, digital trigger, no-trigger timeout, generator loopback) need the standard fixture below and are reported as skipped, with the reason, when the wiring is not declared as that fixture. Their tolerances are, for instance, 1 kHz +-2 %, top 2.8 V +-12 %, base 0 V +-0.15 V, duty 50 +-3 %.
| Input | Standard fixture |
|---|---|
| A0 | the scope's own signal-generator output (MCX, 0-3 V) |
| A1 (1x probe), A2 (10x probe), A3 (1x probe) | the scope's own 1 kHz calibration output (~2.8-3 V square, slow ~8 us RC rising edge, fast ~0.3 us falling edge) |
| D0, D2, D5, D7 | the same calibration output |
| D1, D3, D4, D6 | left open |
Declare the wiring once, in state/fixture.json of the data folder, or for a single run with --fixture standard:
python -c "from openm454 import fixture; fixture.save('standard', 'standard')"
python -m openm454 selftest # uses the recorded wiring
python -m openm454 selftest --fixture standard # the fixture is wired right now (this run only)
fixture.save(analog='standard', digital='standard', note=None) writes {"analog": ..., "digital": ..., "since": ...};
"standard" for a group of leads means the standard fixture (any other value means something else is connected and
the signal checks on those leads are skipped). A missing file means unknown wiring, so a fresh install runs only the
wiring-free checks; so does a file that cannot be read (the result's note says why; a UTF-8 byte-order mark is
fine). Update the file whenever the wiring changes.
The generator is used only with the analog fixture and without --no-awg: V09 parks it at 0 V for the no-trigger
timeout, and V10 drives a 2 Vpp sine and a 0-3 V square into A0, then turns it off (also when one of its captures
fails). --no-awg (and the MCP default) leaves the generator untouched; V10 is then left out of the results, and V09
is skipped when the last awg setting left the generator running (run awg off first to include it). --quick runs
fewer repetitions (about half the time). The result lists every check with measured and expected; the exit code
is 1 when a gating check failed.
Which command for what
| Command | Use it for |
|---|---|
capture |
one record around an event or trigger, e.g. catch a brownout: --trig A1:fall:3.0V --trig-mode normal --timeout 1h |
watch |
repeated triggered snapshots with measurements (frequency, Vpp, ...) every N seconds |
stream |
untriggered recording with live statistics, analog and digital at their own rates, decoding the record |
log |
continuous gap-free recording or monitoring over minutes to hours: min / max / mean per interval, nothing missed between lines; with --hook, your own Python code sees every sample as it arrives and can raise events or stop the run (see Live processing hook) |
Monitoring with log
log prints one statistics line per --every of signal time (default 1 s), so a long run is a table of
vmin_v / vmax_v / vmean_v per interval: dips and spikes show in vmin_v / vmax_v even when the mean hides
them. A 3.3 V rail for 8 hours, one line per ~10 s, nothing saved (--storage none):
python -m openm454 log -c A1 --vdiv 0.5V --offset 2.1V --rate 10k --duration 8h --every 10s --storage none
The window is ~0.09..4.09 V and one ADC step is ~17 mV (pick V/div as described under
Limits and safety). Without --storage none every sample is kept in RAM (the default) and
written at the end: 72 MB for 2 h at 10 kSa/s, 288 MB for 8 h. Ripple on a rail: AC coupling and a small V/div,
which loses the DC level:
python -m openm454 log -c A1 --coupling ac --vdiv 20mV --rate 1M --duration 1h --every 1s --storage none
Check a plan first with log --plan (same options, no device). Over MCP: log_plan, then record_start with
mode: "log" (it takes vdiv, offset and coupling too), record_status (the last lines) and record_stop. A
recording started over MCP is stopped when the MCP client exits, so an unattended multi-hour log needs the Claude
session to stay open, or the CLI run in the background (openm454 log ..., ended with openm454 stop).
How fast can log refresh?
--every can be as small as one 64-sample block, and lines are cut at exact signal-time boundaries (--every rounded to whole 64-sample blocks, see plan.every_s), not at USB transfers. Measured
on the M454 at 150 MB/s (A0-A1 + D0-D7 at 50 MSa/s, --storage none, 10 s each):
--every |
live lines | arrival gaps at the reader | scope backlog |
|---|---|---|---|
| 1 s | 1/s | ~0.94 s | 1.4 MB, flat |
| 100 ms | 10/s | ~103 ms | 1.4 MB, flat |
| 10 ms | 100/s | ~13 ms | 1.3 MB, flat |
| 1 ms | 1001/s | bursts, at most 8.6 ms apart | 1.2 MB, flat |
Printing 1000 lines/s while recording 150 MB/s is therefore not the limit (the lines carry statistics, not the samples). The data reaches the PC in USB transfers of ~1 MB (about 7 ms at
150 MB/s), so lines arrive in bursts and the earliest reaction is ~10 ms after the signal. The real limit is the
reader: an agent reading JSON lines or polling MCP record_status reacts in seconds at best. To react within
milliseconds, run your own code next to the data with a hook.
Live lines of log and stream
log: when this channel set's readout is measured first (above 150 MB/s and not yet known, or --remeasure), a
probe line; then a line with "started": true, the plan and out_stem; then one line per --every, then the
result. With --hook, event lines ({"event": ...}, no i) are interleaved as they happen; see Live processing hook. Lines are cut at whole 64-sample blocks, so a line covers --every rounded to a multiple of 64 / rate (at
10 kSa/s --every 10s gives 9.9968 s); plan.every_s shows the value used. A live line has:
| Key | Meaning |
|---|---|
i |
line number from 0 |
t_signal_s |
signal time since the start of the recording, at the end of this line (samples / rate) |
t_wall_s |
PC clock seconds since the run started, when this data arrived (t_wall_s - t_signal_s is the lag) |
backlog_mb |
MB the scope has acquired but the PC has not read yet (an estimate); the run stops near plan.backlog_stop_mb (~503 MB): HOST_TOO_SLOW, or ring_full for a --force log planned to end there |
backlog_growth_mb_s, ring_full_in_s, gapfree_projected_s |
only while the backlog grows: MB/s over the last ~2 s, seconds until the guard stops the run, gap-free signal seconds the run will hold |
A0..A3 |
vmin_v, vmax_v (whole ADC codes, LSB = V/div / 29.375) and vmean_v: volts at the probe tip over this line's interval only |
D0..D7 |
high_frac (0..1, share of samples high), edges (transitions in this interval) |
chunk_saved |
disk storage: index N of the chunk_NNNNN.npz finished during this line |
ram_kept_s |
RAM ring (--keep-last, forever): seconds of signal now held |
final |
true on the last, possibly partial, line |
hook_ms |
with --hook: milliseconds spent in the hook since the previous line |
stream: lines have i, t_signal_s, t_wall_s (seconds since this record's start; it restarts at every record of
a chained or forever stream, while t_signal_s runs over the whole stream), A0..A3 with vmin_v, vmax_v,
vmean_v and n (samples in this line), and D0..D7 with high_frac, edges. A line is emitted once signal
time passes the next --every boundary, so it is cut at USB event boundaries. A single record ends with
final: true; a chained stream adds record: k to every line and record_end: true at the end of a record. There
are no backlog fields. The final result has the full measurements (--measure, --window, --edges, --dump), files,
hints and warnings; chained streams add records, record_s, signal_s, seams_s, seam_max_s, stopped_by.
Live processing hook (log --hook)
A hook is a Python file of your own that sees every sample of a gap-free log recording as it arrives and can react
at once: a threshold alarm, an edge counter, a trigger for other equipment, an early stop. It is the way for an agent
to act on real-time data faster than it can read lines (see Monitoring with log).
python -m openm454 log -c A1,D0 --rate 1M --duration forever --storage none --hook myhook.py
The file is loaded before the scope is opened. Loading imports the file, so its top-level code runs. A missing file, a
syntax or import error, a sys.exit(), or a missing process() gives INVALID_ARG (exit 2) and nothing is recorded.
With --plan the file is loaded and checked the same way; no hook function is called and the scope is not touched.
The file runs as a registered module (dataclasses and the like work), its folder is added at the end of the import
path (helper modules next to it import), and print() in it goes to stderr, so stdout stays one JSON document per
line.
The file
def setup(info): # optional, once before the first sample
...
def process(block): # required: called for every piece of the recording, in order
return None # or a dict: an event
def finish(summary): # optional, once at the end; a returned dict goes into the final result
...
info has rate_hz, analog, digital, scale (per analog channel: volts_per_code, zero_code, probe, vdiv_v),
every_s, duration_s, storage, block_samples, piece_s (the planned piece size in seconds). summary of finish() has
samples, duration_s, stop_reason, calls, events.
Every sample is passed exactly once, in order, without gaps, in pieces of about 10 ms of signal (shorter when --every is shorter, since a piece never spans more than one line;
longer when a USB transfer holds more, e.g. ~20 ms roll packets at slow rates; info['piece_s'] is the planned size). Lines, statistics and storage are unchanged by the hook.
The block
| Member | Meaning |
|---|---|
t0_s, t1_s |
signal seconds since the start at the beginning and end of the piece |
rate_hz, n |
sample rate and samples per channel in this piece |
analog, digital |
channel lists |
codes(ch) |
raw ADC codes (uint8) of an analog channel |
volts(ch) |
volts at the probe tip (float32) |
bits(ch) |
one uint8 0/1 per sample of a digital channel |
packed(ch) |
digital bits packed 8 per byte, little-endian |
times() |
float64 signal time of every sample |
The arrays are the recording's own and read-only: an in-place edit (c -= 128) raises and ends the run with
HOOK_ERROR, the stored data unchanged. Compute on a copy (c.astype(np.int16) - 128). Pieces vary in length (n):
carry what you need across calls (the last sample, a running state) or an edge or crossing at a piece boundary is
missed.
Events and stopping
process() returns None or a dict. A dict is an event and is printed at once as
{"event": {...}, "k": n, "t_signal_s": ..., "lag_s": ..., "t_wall_s": ...}: t_signal_s is the end of the piece,
lag_s how far behind live the hook was (seconds of newer signal the scope had already taken), k the event number
from 0, t_wall_s PC seconds since the run started. Put "stop": true in
the dict to end the recording cleanly (acq.stop_reason hook, ok true): the recording ends with that piece, and
nothing after it is kept or counted (in a RAM ring the history before the event stays). Any other return type is a
HOOK_ERROR.
The dict is printed as JSON: numpy scalars and arrays in it become plain numbers and lists, NaN and infinity become
null; anything else JSON cannot hold (a set, an object) is a HOOK_ERROR. The same holds for the dict finish()
returns.
Live lines get hook_ms (time spent in the hook since the last line). The final result gets a hook block: file,
calls, events, seconds, load (hook seconds per signal second), lag_max_s, stopped_run, last_event and
result (the dict finish() returned, if any). A load above 1 or a lag_max_s above 1 s adds a warning that the
hook fell behind.
Hook errors
Any exception raised by hook code, sys.exit() and KeyboardInterrupt included, is a HOOK_ERROR. An exception in
process() ends the run with HOOK_ERROR (exit 2; error.trace has the traceback tail;
acq.stop_reason is hook_error). The data recorded up to and including that piece is kept and saved, and finish() is
not called. An exception in finish() also gives HOOK_ERROR with error.trace, but the recording is complete and
acq.stop_reason stays what ended it. An exception in setup() gives HOOK_ERROR (the message names the exception,
there is no error.trace) before anything is recorded. When the run ends with a device error or Ctrl+C, finish() is
still called and the error result carries the hook block.
Hook performance
The hook runs in the thread that decodes the data, so it has to keep up. If it falls behind, the queue grows and the
host-speed guard stops the run with HOST_TOO_SLOW at 768 MiB (~805 MB) queued (data kept). Measured at 150 MB/s: a light hook
(count D0 edges with bits(), min/max of A1 codes, an event every 100 ms, stop at 2 s) made 144 calls in 2 s with
load 0.022, lag_s median 0.2 ms and max 7 ms, and stopped the run itself at 2.0 s. A heavy hook (volts() of both
analog channels plus an FFT per piece) had load 3.16, lag up to 2.2 s and a warning. So: use numpy on codes() /
packed() instead of volts() / bits() for every sample, keep the work per piece small, accumulate across pieces
in module-level state, and lower --rate or use fewer channels if needed.
Hook safety
The hook must not talk to the scope: it is busy recording, so no openm454 commands from the hook. Files, sockets,
serial ports and subprocesses are fine. A hook is your own code running with your rights: only run hook files you
wrote or trust.
Hook over MCP and from Python
record_start takes hook (the absolute path of the file; a relative path is refused because the recording runs in
the MCP server's folder) with mode: "log"; with mode: "stream" it is INVALID_ARG. The hook's imports must be
installed in the Python that runs the MCP server.
record_status then adds events (the last lines event lines, default 3, at most 20) and events_seen (event lines in
the last 256 KiB of the output; the total is hook.events in the result of record_stop).
From Python: logger.run(..., emit=print_or_collect, hook=openm454.hook.Hook(path)); the event lines go to emit.
Example: power-rail dip alarm
A1 on a 3.3 V rail; the hook reports the first sample below 3.0 V in each piece, comparing raw codes against a precomputed integer limit, so no volts array is built per piece:
import numpy as np
LIMIT_V = 3.0
def setup(info):
s = info['scale']['A1']
global LIMIT_CODE
LIMIT_CODE = int(np.ceil(s['zero_code'] + LIMIT_V / (s['volts_per_code'] * s['probe'])))
def process(block):
c = block.codes('A1')
i = np.flatnonzero(c < LIMIT_CODE)
if i.size:
return {'dip_at_s': round(block.t0_s + i[0] / block.rate_hz, 6), 'min_v': round(float(block.volts('A1').min()), 4)}
python -m openm454 log -c A1 --vdiv 0.5V --offset 2.1V --rate 1M --duration forever --storage none --hook dip.py
Run on the M454 with A1 on the calibration square, which is below 3.0 V most of the time: one event per ~20 ms
piece at 1 MSa/s, load 0.005. The light and heavy hooks under
Hook performance were measured on the M454 too.
Limits and safety
From the vendor datasheet:
| Input | Limit |
|---|---|
| analog BNC | -200 .. +200 V peak (DC + AC); 1 MOhm // ~16 pF; 8 bits, 1 mV/div .. 5 V/div |
| digital leads | -30 .. +30 V at the input end of the lead; 1 MOhm // ~16 pF; threshold 0..5 V (--threshold, default 1.4 V) |
| generator (MCX) | 0 .. 3 V |
| calibration output | ~3 V, ~1 kHz square |
- The probe ground is the PC's ground (through USB, and through the PC's mains earth if it has one). Connect it only to a point at the same potential, never to a "hot" ground or any point referenced to the mains.
- The on-screen range is not the safe range. At 1x the screen reaches about +-41.7 V (5 V/div with the offset at its limit), and at 10x about +-417 V, but the BNC limit is +-200 V peak and the probe's own rating applies.
- Running past the screen saturates the ADC; the data of that part is clipped.
Choosing V/div and offset for a DC rail with DC coupling: the screen shows 8 divisions, the farthest reachable edge is +8.34 divisions, so use V/div >= V_rail / 8.34 and set the offset so the rail is on screen: with the default offset 0 V a positive rail above ~4 x V/div is off screen (at 1 V/div the window is -4.02..+3.98 V).
| Rail | V/div (1x) | Note |
|---|---|---|
| 3.3 V | 0.5 V | offset 2.0 V: window 0..4.0 V (2.1 V: 0.09..4.09 V; at 2.0 V a collapse to 0 V reads as 0 V); 0.2 V/div cannot reach 3.3 V |
| 5 V | 1 V | offset 2.5 V: about -1.5..6.5 V |
| 12 V | 2 V | offset 6 V: about -2..14 V |
| 24 V | 5 V | offset 12 V: about -8..32 V |
| above ~41 V | use a 10x probe | --probe 10x, limits x10; stay below the probe's rating |
MCP server
openm454-mcp (or python -m openm454.mcp_server) serves the scope over the Model Context Protocol (stdio, no extra
dependencies; it is installed with pip install openm454):
claude mcp add -s user openm454 openm454-mcp
This form works in every shell (PowerShell, cmd, bash). -s user makes the server available in every folder
(without it, only in the folder where you ran the command). If pip's Scripts folder is not on PATH, give the
full path to openm454-mcp. DSWave is found by itself (search order above), or record it once with
python -m openm454 setup --dswave <folder>. Check with claude mcp list. Other clients:
{"mcpServers": {"openm454": {"command": "openm454-mcp"}}}.
An environment variable (OPENM454_DSWAVE, OPENM454_HOME) reaches the server only through the environment Claude
starts it with: give it when registering (-e). Claude starts the server in the background, so a variable set later
in a shell never gets there.
Put the server name before -e (-e takes several values and would swallow a name after it), then --, then the
command:
claude mcp add -s user openm454 -e "OPENM454_DSWAVE=<folder>" -- openm454-mcp
In Windows PowerShell, an npm-installed claude runs through a .ps1 wrapper that drops a bare --, which gives
"missing required argument". Write the separator in quotes ('--'), call claude.cmd, use
"--env=OPENM454_DSWAVE=<folder>" with no separator, or run the command in cmd. Restart Claude afterwards. Each tool call runs one CLI command in a child process,
so the scope is opened and released per call, a hang never blocks the server, and every rule of the CLI applies.
11 tools:
| Tool | Runs | Key arguments (defaults) |
|---|---|---|
status |
info |
none. Connected? model, USB speed, firmware / FPGA state, last generator setting |
capture |
capture |
channels (A0-A3,D0-D7), timebase / duration / rate / samples, trigger (SRC:EDGE[:LEVEL]), trig_mode (normal with a trigger, force without), timeout_s (5), trig_pos, holdoff, trig_retries (4), decode, edges, window, save (true), plot (true), the vertical options (vdiv, offset, probe, coupling, bw20, threshold). Per channel frequency, Vpp, min/max/mean, duty, rise/fall (digital: state, frequency, duty, edges), trigger result, decoded frames, file paths. A trigger that does not fire gives TIMEOUT_NO_TRIGGER |
analyze |
analyze |
file (required), channels, window, edges, decode; measurements of a saved .npz |
plot |
plot |
file (required), channels, window, out; returns the .png path |
awg |
awg set / dc / off / status |
action (required: set, dc, off, status); wave, freq, high, low (or vpp, dc), duty, phase; level for dc. The output keeps running after the call |
log_plan |
log --plan |
rate (required), channels, duration, every, storage, keep_last, force, the vertical options; no device needed |
record_start |
log or stream in the background |
mode (required: log or stream), channels, rate, duration, every, storage (log: ram / disk / none; stream: only none, nothing saved), keep_last, force and hook (log only; a hook file, record_status then returns its events), decode (stream only), the vertical options (vdiv, offset, probe, coupling, bw20, threshold); returns a job id. One recording at a time, the scope is busy meanwhile |
record_status |
none (reads the job's output) | job (the latest), lines (3); latest status lines, and the result once done; with a hook, also events and events_seen |
record_stop |
stop |
job, wait_s (120); stops cleanly and returns the result |
release |
release |
none; frees a scope left running by a crashed program |
selftest |
selftest |
quick (true), fixture (file; or standard), awg (false), detail, max_wall_s (1260) |
Results are summaries; on capture, analyze and selftest, detail: "full" returns the CLI's whole document. The
samples stay in the .npz. Every device call has a hard time limit, after which an error (TIMEOUT_WATCHDOG) is
returned and the CLI's own watchdog stops the scope: capture and selftest take max_wall_s (default for capture:
180 s plus the trigger timeout and the record length; for selftest: 1260 s, above the CLI's own 1200 s watchdog);
status, awg and release are limited to 60 s.
selftest runs selftest --quick (--quick dropped when quick is false), with --no-awg unless awg is true and
--fixture standard when fixture is standard, so by default it never touches the generator and only runs the
signal checks if the wiring is declared. Pass fixture: "standard" only after the user has wired and confirmed the
standard fixture, and awg: true only when the generator output goes to A0 and nothing else. The summary has
passed_checks, failed_checks (measured vs expected), skipped_checks (grouped by reason) and counts that match
those lists (checks not run after a device failure count as skipped), plus fixture, note and seconds. A failed
run returns the error SELFTEST_FAILED, whose hint tells a failed signal check (usually the wiring) from a failed
wiring-free one (the scope or its USB link); a device failure returns that device error (with stopped_at).
CLI commands with no MCP tool:
| Command | Why |
|---|---|
watch |
prints one JSON line per capture over many captures (20 by default, or until --count / --for / --stop-when / openm454 stop), which does not fit one request and one reply; use record_start with mode: "stream", or call capture repeatedly |
setup |
one-time configuration of where DSWave is; run it once in a terminal: every later tool call reads the saved path (state/dswave.json) |
bench |
maintenance: measures this PC's USB readout speed, which log uses |
recover |
resets and reloads the FPGA of a hung scope; kept a deliberate CLI step |
debug |
raw register access, deliberately not exposed to agents |
stop |
record_stop uses it internally |
awg sweep, burst, mod |
the awg tool covers set, dc, off, status only |
An agent that has a shell can use the CLI for all of these.
Errors
Every error has a code, a message and a hint, so the next step is clear:
| Code | Exit | Meaning |
|---|---|---|
NOT_FOUND |
3 | the scope is not connected (USB 2A0E:0080 not seen) |
BUSY |
4 | DSWave or another openm454 process holds the scope |
TIMEOUT_NO_TRIGGER |
5 | nothing triggered within --timeout (--trig-mode normal) |
TRIGGER_OUT_OF_RANGE |
2 | the trigger level is outside the channel's visible range (the message gives it); in normal and auto mode (with --trig-mode force the trigger is not used and the level is clamped) |
FIRMWARE_SOURCE |
6 | DSWave 1.2.1 not found (or the file is not 1.2.1), so the FPGA bitstream cannot be extracted; the message lists what was searched, the hint names openm454 setup --dswave <folder> |
UNSUPPORTED |
6 | the scope is connected at USB 2 speed; capture, watch, stream and log need a USB 3 (SuperSpeed) link |
INVALID_ARG |
2 | an argument is wrong or impossible for this channel set, or the log --hook file cannot be used (missing, syntax or import error, no process()) |
STREAM_BETTER, NEEDS_FORCE |
2 | log longer than its gap-free limit (see above) |
HOOK_ERROR |
2 | log --hook: the hook file raised in setup(), process() or finish(), or process() returned something other than None or a dict; for process() / finish() error.trace has the traceback tail and the data recorded so far is kept; a setup() failure records nothing |
SELFTEST_FAILED |
1 | MCP selftest only (the CLI exits 1 and lists the checks): a check failed; failed_checks has measured vs expected (a failed signal check usually means the wiring differs from the standard fixture, a failed wiring-free check points at the scope or its USB link) |
UNSAFE_OPERATION |
7 | refused by the safety rules |
TIMEOUT_WATCHDOG, TIMEOUT_DATA, USB_ERROR, PROTOCOL_ERROR, HOST_TOO_SLOW |
6 | the device or the host failed during a run; data up to then is kept |
Layout
openm454/driver: USB transport with a safety whitelist, device bring-up (FPGA, security handshake, calibration), register compiler (mso.py, reproduces DSWave's register programs exactly), stream decoder (packet.py), acquisition engine, loop-mode logger (logger.py), live processing hook forlog --hook(hook.py), measurements, plots, AWG, CLI, self-test, protocol decoders (decode.py), synthetic bus traffic for tests (protosim.py) and the MCP server (mcp_server.py).tests/offline tests: golden register and AWG programs, the stream decoder, the acquisition engine against a simulated device, measurements, protocol decoders on synthetic traffic, the MCP server, safety rules (python -m pytest tests).docs/PROTOCOL.mdthe recovered USB protocol..github/workflows/: CI (the offline tests on Windows and Linux, Python 3.10-3.14, withOPENM454_NO_USB=1, which makes the driver refuse to open USB; a check that tracked files carry no local paths; a package build) on pushes to main and on pull requests; a tagvX.Y.Zequal toopenm454.__version__tests, builds and publishes to PyPI (trusted publishing) and creates a GitHub release.
What is NOT here
The vendor application, its embedded FPGA bitstream and firmware images, decompiled vendor code, raw USB traces (they contain the device's security key), the reverse-engineering workspace and the vendor datasheet are kept out of this repository.
Safety
The driver never sends the firmware-upgrade request (bRequest 0xB3), never writes the device EEPROM, and configures
the FPGA only when the device reports it unconfigured or when openm454 recover resets a hung scope (always from
the verified bitstream, as the first write of the session).
License
GPL-3.0 (see LICENSE). DSView (GPLv3) by DreamSourceLab was used as a protocol reference. DSWave and DSCope are
trademarks of DreamSourceLab; this project is not affiliated with DreamSourceLab.
Metadata
Release files for openm454 0.2.0
For a detailed explanation of source distributions (sdists) and built distributions (wheels), please see the package formats documentation.
Source distribution (sdist)
| File | Size | Uploaded | Provenance | Actions |
|---|---|---|---|---|
| openm454-0.2.0.tar.gz | 318.3 kB | With provenance | Details for openm454-0.2.0.tar.gz |
Built distribution (wheel)
| File | Interpreter | ABI | Platform | Actions Reset |
|---|---|---|---|---|
| openm454-0.2.0-py3-none-any.whl | Python 3 | none | any | Details for openm454-0.2.0-py3-none-any.whl |
Total release size: 526.5 kB
Release files / openm454-0.2.0.tar.gz
| Download URL | openm454-0.2.0.tar.gz |
|---|---|
| Size | 318.3 kB |
| Tags | Source |
|
SHA-256 checksum How to use checksums |
fddf1e0ec5000517aa55e342980bfa6197ae4339f15333b1b167557583312481
|
|
BLAKE2b-256 checksum How to use checksums |
4441f8e03a06f9b5dc30d3083fa2659fbfa49f3e2276f62a0e7b8870ed3a3601
|
| Upload date | |
|
Uploaded using Trusted Publishing? What is trusted publishing? |
Yes |
| Uploaded via |
twine/7.0.0 CPython/3.13.16
|
Provenance
Provenance describes where a file came from. On PyPI, provenance is shared via attestations, which provide a verifiable record of the build or publishing details. View details, limitations and caveats.
PyPI Publish Attestation
PyPI verified that this artifact, at this checksum, originated from the publisher listed below.
Signed by GitHub Actions, verified by PyPI on Nov 7, 2026.
Transparency logRelease files / openm454-0.2.0-py3-none-any.whl
| Download URL | openm454-0.2.0-py3-none-any.whl |
|---|---|
| Size | 208.3 kB |
| Tags | Python 3 |
|
SHA-256 checksum How to use checksums |
24af1104995f7e215c0fe831114c44dfbcbe3947ff74b85bc58e1b76960f97a9
|
|
BLAKE2b-256 checksum How to use checksums |
4f383a5782f3a0da60524413dc39df1ce68d3f1ae368a880adb380211778927b
|
| Upload date | |
|
Uploaded using Trusted Publishing? What is trusted publishing? |
Yes |
| Uploaded via |
twine/7.0.0 CPython/3.13.16
|
Provenance
Provenance describes where a file came from. On PyPI, provenance is shared via attestations, which provide a verifiable record of the build or publishing details. View details, limitations and caveats.
PyPI Publish Attestation
PyPI verified that this artifact, at this checksum, originated from the publisher listed below.
Signed by GitHub Actions, verified by PyPI on Nov 7, 2026.
Transparency logSecurity for openm454 0.2.0
Provenance
Provenance describes where a file came from. On PyPI, provenance is shared via attestations, which provide a verifiable record of the build or publishing details. View details, limitations and caveats.
Full provenance
Both files in this release have been uploaded with provenance metadata.
Attestation for openm454-0.2.0.tar.gz
PyPI Publish Attestation
PyPI verified that this artifact, at this checksum, originated from the publisher listed below.
Signed by GitHub Actions, verified by PyPI on Nov 7, 2026.
Transparency logA note on package security
PyPI is a repository for distributing packages, not a service for auditing them. While third-party security researchers and our staff do their best to detect and remove malicious packages, PyPI cannot guarantee the quality, security or suitability of any given package.
To learn more about protecting yourself from malicious actors, visit our package security documentation.