PLESTY Keithley 2410 SourceMeter

A PLESTY device module for the Keithley 2410 SourceMeter, a high-voltage source-measure unit (SMU). An SMU sources a voltage or a current and measures both at the same time. The 2410 sources up to 1100 V or 1.05 A and talks SCPI (the standard text command language of laboratory instruments) over VISA (the standard software layer for GPIB, serial and USB instruments) on GPIB or RS-232.

The source can put hazardous voltage on its terminals, so this module refuses before it acts. Read Safety rules before switching anything on.

This module replaces an older PLEXY driver for the same instrument. It is not backwards-compatible with it.

Supported hardware

Instrument Link Notes
Keithley 2410 SourceMeter GPIB or RS-232 The preflight check requires MODEL 2410 and refuses other models

Installation

git clone https://gitlab.com/plesty/hub/devices/keithley/plesty-k2410-source-meter.git
cd plesty-k2410-source-meter
uv sync

A VISA backend must be available to talk to a real instrument (for example NI-VISA, plus a GPIB adapter for a GPIB link). The mock address needs none.

Configuration

The address of the instrument is configuration, not code. Copy .env.example to .env and edit it:

Variable Default Meaning
K2410_ADDRESS GPIB0::24::INSTR VISA resource name of the instrument
K2410_TCP_PORT 5555 ZMQ port the device server binds for remote clients
  • GPIB: GPIB0::<n>::INSTR. The factory GPIB address is 24; the front panel shows the current one under MENU > COMMUNICATION > GPIB.
  • RS-232: ASRL<n>::INSTR. Set the instrument to 9600 baud, 8N1, CR terminator.
  • mock: the word mock runs an in-memory simulator of the instrument, with no hardware.

Each value resolves as CLI argument, then environment variable.

Quick start

Run the device as a TCP server:

cp .env.example .env   # set K2410_ADDRESS
uv run python -m plesty.k2410_source_meter
uv run python -m plesty.k2410_source_meter --address ASRL3::INSTR --tcp-port 5555
uv run python -m plesty.k2410_source_meter --address mock   # simulator

Use it from Python (local, on the simulator):

from plesty.k2410_source_meter import Device

dev = Device(address="mock")  # a VISA address such as "GPIB0::24::INSTR" for the instrument
dev.init()
dev.connect()
reading = dev.source_and_measure(
    source_function="VOLT", level=5.0, compliance=1e-3, source_range=20.0
)
print(reading["voltage"], reading["current"])
dev.disconnect()

Preflight

Before the server binds its port, the module opens the configured address, reads the identification string (*IDN?) and requires MODEL 2410. When the configured address does not answer as a 2410 but another GPIB address does, the message names that address and tells you to set K2410_ADDRESS to it; the address is never changed for you. --skip-preflight serves anyway, for protocol work.

Safety rules

The 2410 answers a bad setting with an error-queue entry only after it has acted on the neighbouring ones, which is too late for a source that may carry 1100 V. So every rule below is checked in software before the first command is sent; a violation raises a refusal and sends nothing.

  • Explicit compliance. The compliance is the limit on the quantity that is not sourced (amperes while sourcing voltage, volts while sourcing current). output_on is refused until configure_source set one explicitly since connect. There is no default compliance.
  • voltage_limit defaults to 20 V. This is the instrument's hardware over-voltage protection. It is applied as 20 V on connect, so any higher voltage needs a deliberate write of voltage_limit first, while the output is off. A level above it is refused.
  • output_on ramps from 0 V. It switches the output on at 0 and steps up to the configured level (steps of at most 10 V, a tenth of the range, with a short wait between steps). A warning is logged when voltage_limit is above 30 V.
  • Operating boundary. Above 21 V the current compliance must be 21 mA or less; above 21 mA the voltage compliance must be 21 V or less.
  • Compliance floor. A compliance may not be below 0.1 % of the measure range in force.
  • Sourced quantity on the source range. The 2410 measures the quantity it sources on the source range and rejects a measure range for it (error 823). So voltage_measure_range and its _auto flag are refused while sourcing voltage, and the current ones while sourcing current; set voltage_range or current_range instead.
  • No reading with the output off. measure never switches the output on.
  • Stepped level changes. With the output on, set_level walks to the new level in steps.
  • Refusals when unsafe. output_on is also refused while the instrument's error queue is not empty or, when interlock is on, while the interlock line is open.
  • Shutdown on failure. safe_shutdown walks the level to zero in steps and then switches the output off. Every failure with the output on, and disconnect, do the same. A refusal changes nothing, so it does not shut down.

Parameters

Read with dev.query(name) and write with dev.write(name, value); dev.state returns all of them at once. Values are in volts, amperes and seconds.

Parameter Type Range / options Default Meaning
source_function str VOLT, CURR VOLT What is sourced
voltage_level float, V -1100 to 1100 0 Level when sourcing voltage
current_level float, A -1.05 to 1.05 0 Level when sourcing current
voltage_range float, V 0.2 to 1100 20 Voltage source range
voltage_range_auto bool off Instrument picks the range
current_range float, A 1e-6 to 1.05 1e-4 Current source range
current_range_auto bool off Instrument picks the range
current_compliance float, A 1e-9 to 1.05 1e-4 Current limit while sourcing voltage
voltage_compliance float, V 2e-4 to 1100 1 Voltage limit while sourcing current
voltage_limit float, V 20, 40, 100, 200, 300, 400, 500, 1100 20 Over-voltage protection (1100 is none)
sense_function str VOLT_CURR, VOLT, CURR, RES VOLT_CURR What measure returns
current_measure_range float, A 1e-6 to 1.05 1e-4 Current measure range
current_measure_range_auto bool on Instrument picks the range
voltage_measure_range float, V 0.2 to 1100 20 Voltage measure range
voltage_measure_range_auto bool on Instrument picks the range
nplc float, PLC 0.01 to 10 1 Integration time of one reading, in power line cycles
source_delay float, s 0 to 9999.999 0 Settling time before each measurement
remote_sense bool off 4-wire measurement at the load
terminals str FRON, REAR FRON Which connector carries the output
output_off_mode str ZERO, NORM, GUAR, HIMP ZERO Terminal behaviour while off
interlock bool off Hardware interlock of the rear Digital I/O connector
output_state bool read only off Whether the output is on

The compliance written by configure_source is what allows output_on; the value applied on connect is a conservative baseline, not a choice. Change source_function, remote_sense, terminals and voltage_limit only while the output is off.

Functions

Function Arguments Returns
get_identity() manufacturer, model, serial_number, firmware
configure_source(source_function, level, compliance, source_range=None) compliance is required values read back; output stays off
output_on() output state
output_off() output state; immediate, no step down
safe_shutdown(step=None, delay=None) output_state, level
set_level(level, step=None, delay=None) level read back
measure() voltage, current, resistance, timestamp, status, compliance, overrange, ovp_tripped
sweep(source_function, start, stop, points, compliance, spacing="LIN", delay=None, return_to_zero=True) 2 to 2500 points iv_curve, points, compliance_hit
read_errors() errors, count (reads and clears the queue)

A compliance flag of True on a reading means the source did not reach its setpoint, so the reading is a limit and not a measurement.

Composite operations

Two methods of Device chain the functions above into a complete measurement. Both step the source down and switch the output off afterwards, also on failure.

  • source_and_measure(source_function, level, compliance, source_range=None, keep_output_on=False) configures, switches on, takes one reading and switches off.
  • run_iv_sweep(source_function, start, stop, points, compliance, spacing="LIN", delay=None, source_range=None) configures at the first level, switches on, sweeps and switches off.

The full generated reference (types, ranges, per-function arguments) is in the documentation under docs/; build it locally with CI= uv run plesty check.

Open items for the field test

Verified on a real 2410 over GPIB (2026-10-20, open terminals, at most 5 V):

  • The instrument accepts :SYST:ERR:ALL? (read the whole error queue at once).
  • :OUTP:INT:TRIP? answers 0 while the interlock is open and 1 while it is closed. The manual contradicts itself here; its SCPI reference table is right. With the interlock enabled and open, :OUTP ON is blocked with error 802.
  • The five-element :READ? reading and its status bits; output_on stepping up from 0, set_level, safe_shutdown and a 6-point sweep; GPIB terminators.

Still to confirm, marked in the code:

  • Whether NONE is the right setting for the 1100 V over-voltage protection.
  • Whether :READ? behaves as expected with the output off (the driver refuses it anyway).
  • The ramp step size (10 V) and wait (30 ms) on a real load, and compliance on a load.
  • How the integration time is shared between the voltage, current and resistance functions.
  • RS-232 line terminators.

License

LGPL-3.0-or-later. See LICENSE and COPYING.

Metadata

Release files for plesty-k2410-source-meter 0.1.0

For a detailed explanation of source distributions (sdists) and built distributions (wheels), please see the package formats documentation.

Source distribution (sdist)

Source distribution for plesty-k2410-source-meter 0.1.0
File Size Uploaded Provenance Actions
plesty_k2410_source_meter-0.1.0.tar.gz 230.5 kB Without provenance Details for plesty_k2410_source_meter-0.1.0.tar.gz

Built distribution (wheel)

Table of built distributions (wheels) for plesty-k2410-source-meter 0.1.0
File Interpreter ABI Platform Actions
plesty_k2410_source_meter-0.1.0-py3-none-any.whl Python 3 none any Details for plesty_k2410_source_meter-0.1.0-py3-none-any.whl

Total release size: 301.0 kB

Release files / plesty_k2410_source_meter-0.1.0.tar.gz

Download URL plesty_k2410_source_meter-0.1.0.tar.gz
Size 230.5 kB
Tags Source
SHA-256 checksum
How to use checksums
d3899ac899618eb89c6260c6294e400a48368d18288fb2451a8e1b0e5112aac3
BLAKE2b-256 checksum
How to use checksums
5c2d3649beb6ff5213d6170a9f2e41dab89bf4934b0566a07f8f57203290bba0
Upload date
Uploaded using Trusted Publishing?
What is trusted publishing?
Yes
Uploaded via uv/0.9.30 {"installer":{"name":"uv","version":"0.9.30","subcommand":["publish"]},"python":null,"implementation":{"name":null,"version":null},"distro":{"name":"Debian GNU/Linux","version":"12","id":"bookworm","libc":null},"system":{"name":null,"release":null},"cpu":null,"openssl_version":null,"setuptools_version":null,"rustc_version":null,"ci":true}

Release files / plesty_k2410_source_meter-0.1.0-py3-none-any.whl

Download URL plesty_k2410_source_meter-0.1.0-py3-none-any.whl
Size 70.5 kB
Tags Python 3
SHA-256 checksum
How to use checksums
1b1b0cb96b1e1149eceff859bc47ee0854d21266a1b73031ebf2c14aed6df6d0
BLAKE2b-256 checksum
How to use checksums
7ad5c6a1c2e69a01d8b3ea5c27c3559fbb4a7cd6866fdcfe2408cdeab31d358a
Upload date
Uploaded using Trusted Publishing?
What is trusted publishing?
Yes
Uploaded via uv/0.9.30 {"installer":{"name":"uv","version":"0.9.30","subcommand":["publish"]},"python":null,"implementation":{"name":null,"version":null},"distro":{"name":"Debian GNU/Linux","version":"12","id":"bookworm","libc":null},"system":{"name":null,"release":null},"cpu":null,"openssl_version":null,"setuptools_version":null,"rustc_version":null,"ci":true}

Security for plesty-k2410-source-meter 0.1.0

Provenance

Provenance describes where a file came from. On PyPI, provenance is shared via attestations, which provide a verifiable record of the build or publishing details. View details, limitations and caveats.

No provenance found

None of the files in this release were uploaded with provenance metadata.

A note on package security

PyPI is a repository for distributing packages, not a service for auditing them. While third-party security researchers and our staff do their best to detect and remove malicious packages, PyPI cannot guarantee the quality, security or suitability of any given package.

To learn more about protecting yourself from malicious actors, visit our package security documentation.

Release history Release notifications | RSS feed

This release

0.1.0 This release

2 release files No provenance

Anthropic, PBC Visionary sponsor Bloomberg Visionary sponsor Hudson River Trading Visionary sponsor Meta Visionary sponsor NVIDIA Visionary sponsor Microsoft Sustainability sponsor Depot Continuous Integration AWS Cloud computing and Security Sponsor Datadog Monitoring Fastly CDN Google Download Analytics Sentry Error logging StatusPage Status page