RespMech — respiratory mechanics, work of breathing and diaphragm EMG

(c) Copyright 2019–2026 Emil Ingerslev Walsted (emilwalsted@gmail.com), ORCID 0000-0002-6640-7175

DOI

RespMech analyses a time series of respiratory physiological recordings (e.g. exported from LabChart) breath by breath, and calculates:

  • respiratory mechanics
  • inspiratory and expiratory work of breathing (Campbell diagram)
  • diaphragm EMG — RMS envelope, ECG-artefact removal, spectral noise reduction
  • Sample Entropy1,2 (e.g. of diaphragm EMG)

Version 2 is a desktop application with a guided setup, a live preview/QC screen, and a batch runner that opens as a drawer right where you're looking. The physiology is a faithful port of the original v1 code and is locked by golden/characterisation tests — see Correctness.

RespMech — Setup


Install

Recommended — the desktop app. Download the installer for your platform from the latest release: a .dmg (macOS) or .msi (Windows). It bundles its own Python — nothing else to install.

Python package (CLI + core; the GUI is an extra):

pip install respmech            # command-line tool + analysis engine
pip install "respmech[gui]"     # + the PySide6 desktop app (respmech-gui)

From source (developers, or to use the CLI):

git clone https://github.com/emilwalsted/respmech.git
cd respmech
pip install -e ".[dev,gui]"      # gui = PySide6 desktop app;  dev adds the test stack
respmech-gui                     # launch the desktop app

Extras: gui (desktop app), emg (librosa — spectral noise reduction), plots (matplotlib diagnostic figures), dev (tests). See docs/INSTALL.md for details.

See CHANGELOG.md for what changed in each release.

Using the app

respmech-gui

First time in? Choose Explore with sample data on the start screen to load a ready-made synthetic recording — the one shown in the figures below, complete with a heartbeat artefact on the EMG and a little volume drift, so every step has something to demonstrate.

Two tabs:

Tab What you do
Setup Point at your recordings, map the data columns to channels — RespMech suggests roles from the file's own column names, or assign them visually from the data if you prefer — and choose what to save. Settings validate as you type — the status bar flags anything inconsistent.
Preview & QC See the analysis on one file before running the batch: breath segmentation, the Campbell loop and the per-breath table, and dedicated subtabs to tune EMG – ECG reduction and EMG – noise reduction against the live signal. Click a breath to exclude it.

Run & results is a drawer, not a third tab: a compact "Run & results ▸" bar sits below Preview & QC's file list on every subtab, and opens into the batch's progress, per-file status, averaged metrics and output folder — it opens itself the moment you start a run, so you never have to remember to check it.

RespMech — Preview & QC, with Run & results open

Settings are stored as a declarative TOML analysis file (no longer an executable .py). Open, save and switch between analyses — including your recently opened files — from the Analysis menu in the header, which is available on every screen; RespMech marks unsaved edits in the title bar and asks before discarding them. The same actions, plus keyboard shortcuts, live in the window's File menu, alongside View (jump to a tab) and Help (documentation, website, About).

Command line

respmech run settings.toml            # process a batch  (--dry-run computes without writing)
respmech breaths settings.toml [FILE] # list every detected breath/segment (one file, or all matched)
respmech validate settings.toml       # check the settings and the input files
respmech migrate old_settings.py -o settings.toml   # convert a v1 settings file (runs no v1 code)
respmech init new_settings.toml --signals flow,poes,emg [--folder DIR --files MASK --fs 1000]

breaths numbers every breath (or, for an EMG-only signal set, segment) respmech run would build for the given file(s), with its onset, duration, kind and whether it is excluded from the tidal average — the same numbers a real run's output uses, so you can check them before running. It also suggests which untyped breath looks like a forced vital capacity manoeuvre (the longest untyped expiration in the file) and prints a ready-to-paste [[processing.breath_types]] snippet for it.

validate also reports any reference/subject link that will not work once the batch's real file list is known (a reference source not among the analysed files, a linked breath that is excluded or mistyped, a group key matching no file) — advisory unless processing.lung_volume.require_references is set and a reference source is genuinely missing, which fails validation.

run --dry-run prints each file's resolved inspiratory-capacity and forced-vital- capacity reference (own when the reference is the file's own typed breath, e.g. IC ref: own #4 · FVC ref: own #7) whenever this analysis names an IC reference anywhere, or no IC reference — lung-volume columns NaN for a file whose own reference did not resolve.

migrate prints a report of every field moved, renamed, defaulted (not present in the legacy file — e.g. an absent Pgas/Pdi channel, or the derived analysis.signals, which the legacy format never had at all) or dropped.

init writes a commented starting TOML file with only the [input.channels] entries your chosen signal set needs — e.g. --signals flow,poes omits Pgas/Pdi entirely. --folder/--files/--fs are written as real values when given, or left as commented placeholders you fill in by hand; respmech validate (see above) tells you what, if anything, is still missing. respmech validate and the run report both name the signal set an analysis actually uses and what it computes, e.g.:

Signals: flow, poes (derived) · Entropy: 3 columns · Analyses: Breath timing, Work of breathing, Sample entropy · off: Pgas/Pdi (not in signal set)

Signal sets

Not every recording has every channel, so an analysis declares which signals it uses. A new analysis starts with that choice, and everything after it (the channels Setup asks for, the panels Preview & QC shows, the columns and figures a run writes) follows it:

RespMech — choosing the signals an analysis uses

Signal set What you get
Flow only Breath timing, tidal volume and ventilation from flow (and volume). Preview shows a flow–volume loop where the Campbell diagram would be.
Flow + Poes Adds work of breathing and the oesophageal-pressure descriptors.
Flow + Poes + Pgas + Pdi The complete set, and the default: adds gastric and transdiaphragmatic pressure.
… also EMG Any of the three above, plus the diaphragm-EMG analyses.
EMG only No flow channel at all: see EMG-only analyses.

Sample entropy is available in every set once a column is assigned to it. A channel whose role leaves the set is cleared when you change the set, never left assigned to something the analysis no longer declares. The choice is stored as an optional [analysis] table (signals = ["flow", "poes"]); left out, the set is worked out from whichever channels are assigned, so an analysis written before signal sets existed behaves exactly as it did. A run's output only contains the columns of the signals it used, and respmech validate, the run report and each workbook's Provenance sheet name the set and what it computes. Explore with sample data in the File menu opens a reduced sample recording when the current analysis is Flow only or Flow + Poes (without EMG), an EMG-only sample when it is EMG only, and the complete one otherwise.

Breath types and reference manoeuvres

A plain click on a breath only marks it (green on the plots, in the table and in the loop diagram; click again to unmark). Not every breath in a file is tidal breathing. Right-click a breath in Preview & QC ▸ Mechanics (or Ctrl+left-click) and give it a type: Tidal, Excluded, IC manoeuvre, FVC manoeuvre, IC + FVC, Maximal inspiratory effort, Sniff or Other…. A typed breath is left out of the tidal averages exactly like an excluded one, and is measured separately in a Manoeuvres table below the per-breath table and in its own sheet of the file's workbook:

RespMech — a breath typed as an IC manoeuvre, with the Manoeuvres table

  • An IC manoeuvre reports its inspiratory capacity (peak volume above the end-expiratory level of the tidal breaths just before it), timing, peak inspiratory flow and pressure swings, with quality flags for a low effort, an unstable baseline, a measurement that does not repeat and a manoeuvre at the edge of the recording.
  • An FVC manoeuvre reports FVC, FEV₁ and FEV₁/FVC, PEF, the back-extrapolated volume and whether the forced expiration reached a genuine end (ATS/ERS 2019).
  • A maximal inspiratory effort or sniff reports its own peak pressures and EMG, the reference normalisation uses.

The same choices are also available as controls on the toolbar row next to Refresh (a type selector, Use as IC reference for and Reference manoeuvres…). They act on the marked breath and stay disabled until one is marked.

A dedicated recording in which every breath is typed (a separate IC or FVC file) is analysed too; its Manoeuvres table stands in for the breath-by-breath data.

Where a reference comes from. A file's IC, FVC, baseline-IC or maximal-effort reference can be a typed breath in the same file or in a different one, so one IC recording can serve all of a participant's exercise files. Right-click a breath typed as an IC manoeuvre and choose Use as IC reference for ▸ this file / all files of its group / all files, or open Reference manoeuvres… (also on a file's row in the file rail) for the full picker. In the settings file the same choices are [[processing.references]] (one file) and [[processing.reference_defaults]] (a group). The order is always: an explicit entry for the file, then its group's entry, then the file's own typed breath. Per-participant spirometry ([[input.subjects]]: TLC, VC, RV, FEV₁, MVV) applies to every file of that participant, keyed like the cohort summary's groups (the leading filename token, or output.group_regex). An unresolved link is never a hard error: it is a caution in respmech validate, blank values and a note in the run report, unless processing.lung_volume.require_references is set.

respmech breaths settings.toml [FILE] lists every breath with its number, onset, duration and type, and suggests the untyped breath with the longest expiration as a forced vital capacity candidate, with a ready-to-paste [[processing.breath_types]] snippet.

If the automatic detector splits one breath in two (a flow wobble) or merges two (a flat expiration), repair it by hand in Preview & QC ▸ Mechanics: click a trace to cut a new boundary there, or click an existing boundary to join it away (switch on Place separators first, or a click excludes the breath). This corrects the automatic segmentation; an analysis with no repair is unaffected.

EMG-only analyses

A recording with no flow channel, such as a maximal inspiratory or expiratory manoeuvre recorded on the EMG alone, has no inspiration or expiration to find. Choose EMG only in the signal-set picker and say how each recording is cut into segments, the units RMS, integrated EMG and sample entropy are computed on:

  • One maximal manoeuvre: the whole file is one segment.
  • Several efforts or breaths: mark the boundaries yourself in Preview & QC ▸ EMG – segments (click to place, click near a separator to remove it; exclusions and types follow a segment through a renumbering).
  • Tidal breathing — detect bursts automatically: each file is cut at the onset of every burst of inspiratory EMG activity, found on the envelope of the ECG-removed signal (a threshold with hysteresis and a minimum burst and gap length, after Hodges & Bui 1996). Each segment then reports the neural timing of its own cycle (ti_emg, te_emg, ttot_emg, bf_emg), and a recording without a clear burst fails with a named error instead of being cut into noise. In the settings file, method = "fixed_windows" cuts equal windows instead.

RespMech — Preview & QC ▸ EMG – segments, an EMG-only analysis split by separators

The burst thresholds are starting values that have been checked on synthetic recordings only: look at the shaded segments before trusting them on real ones. The noise reference for spectral noise reduction can be the periods between the bursts, or a segment you typed Rest. With whole_file, each EMG channel also reports where the file's peak fell and the mean of its three highest envelope values, a steadier peak level than the single maximum (which can land on a heartbeat when ECG removal is off). Explore with sample data opens an EMG-only sample, already split into its own breaths, when the current analysis is EMG only.

Operating lung volumes, EFL and PEEPi

Typed manoeuvres feed four optional analyses. All are off or blank until their inputs exist, and they only add columns.

Operating lung volumes. Once a file has a resolved IC reference, every tidal breath reports its operating IC, EELV, EILV and inspiratory reserve volume, as a volume above residual volume (from the participant's vital capacity in [[input.subjects]]) and, when a TLC is entered, as an absolute value as well. The IC is by default held constant across the file (the usual reporting convention). processing.lung_volume.ic.eelv_tracking = "within_file" lets it move with the breath's own end-expiratory volume, which shows dynamic hyperinflation; it needs the IC reference to be in the same file, and is left blank when volume trend correction is on. The change in IC and EELV against a baseline recording is reported per file. The run report has a LUNG VOLUMES block naming the datum used.

MFVL, expiratory flow limitation and ventilatory capacity. A file with a typed FVC breath places every tidal flow–volume loop inside that file's own maximal flow–volume curve: how much of the tidal expiration reaches the flow the maximal curve allows (expiratory flow limitation, as a percentage and a yes/no), the fastest the breath could have been exhaled, and the ventilatory capacity and breathing reserve that implies (against a supplied MVV or FEV₁ × 40). This placement uses the file's own typed breaths only (a reference in another file is not used for it), and without an IC manoeuvre in the same file the placement-dependent columns are blank. Preview & QC shows the figure whenever the previewed file has a typed FVC breath, and a run writes flow-volume (tidal in MFVL).pdf.

PEEPi and the modified Campbell diagram. processing.pressure.peepi measures intrinsic PEEP for every breath from the oesophageal-pressure deflection before inspiratory flow starts, and reports the extra work that threshold adds in new columns beside the existing ones (peepi_dyn, wob_in_thr, wobtotal_thr and the pressure–time products including it), drawn as a hatched rectangle on the Campbell diagram. The four detection thresholds are starting values that have not yet been measured on real recordings: treat small deflections with caution.

Normalisation to a maximal manoeuvre. processing.pressure.normalization expresses tidal inspiratory oesophageal and transdiaphragmatic pressure, and EMG, as a percentage of the same person's typed maximal inspiratory effort or sniff, and adds the tension–time indices tt_es and tt_di and, with EMG, the neural respiratory drive index nrdi, on a "Pressure normalised" sheet. The run report and Provenance sheet name the reference, and whether it was a sniff or a maximal inspiration, since those give different diaphragm pressures.

Each of these is described formula by formula, with its known limits, in docs/REVERSE_ENGINEERING.md (§5.11–5.17, all marked v2-only, with no counterpart in RespMech 1.x). A commented example of every settings table is in examples/settings.toml.


Data recording requirements

Input data do not need to be a specific length, but because some outputs are per-time (e.g. minute ventilation) you must specify the sampling frequency of the recording.

The code analyses data breath by breath, and it is imperative that the recording starts with the last part of an expiration and ends with the first part of an inspiration. The recording is trimmed automatically to start at exactly the first inspiration and end at exactly the last expiration.

Breaths are segmented by joining an inspiration with the following expiration, using the flow signal to find the transition. A breath-separation buffer absorbs "wobbly" flow around zero (common in quiet breathing); its length depends on your sampling and breathing frequency. In Preview & QC you can click any breath — e.g. an IC manoeuvre or a cough — to drop it from the analysis:

Breath segmentation and exclusion

Flow and volume conventions. The analysis assumes flow is negative on inspiration and positive on expiration — invert it in Setup if your recording is the other way around. Volume must be inspired volume; it can be inverted, or integrated from the flow signal if your recording has no volume channel. Volume drift (common when integrating from flow) is corrected automatically, with an optional trend adjustment on top — each breath's end-expiratory volume should return to the same baseline, and when it creeps the correction pulls it back. The trend adjustment anchors on the end-expiratory trough of each breath; which troughs count is judged relative to each recording's own volume range, so it needs no tuning at any tidal volume (Preview & QC → Mechanics → Advanced… if a recording does need it):

Volume drift correction

Supported input formats: MATLAB, Excel, CSV/text. (MATLAB files exported from the Windows and macOS versions of LabChart differ — pick the variant in Setup ▸ Advanced.)

Work of breathing

Two options: calculate WOB from each breath's Campbell diagram and average the results, or first build an averaged pressure/volume loop and calculate WOB from that. The two give similar results, but with irregular breaths the averaged loop is more robust. The number of resampling points used when averaging the loop is configurable (a good default is the sampling frequency ÷ 8–10; it must be lower than the shortest inspiration or expiration in the file).

Campbell / PV loop

The Campbell diagram shows the inspiratory work of breathing: the faint loops are the individual breaths, the bold one their average, the diagonal the elastic-recoil line joining end-expiration to end-inspiration, and the shaded triangle the elastic component; the resistive component is the bulge of the trace away from that line. The loop's own enclosed area is not the work of breathing.

Diaphragm EMG

When EMG channels are present, each is conditioned in steps before its RMS envelope and (optionally) sample entropy are measured. The heartbeat (ECG) R-wave is typically several times the EMG amplitude, so if left in it dominates the signal and badly inflates the RMS — it is detected on the clearest channel and subtracted first. Then spectral noise reduction — trained on a diaphragm-quiet reference — cleans the residual noise floor while preserving the inspiratory burst. Both steps are tuned against the live signal on the Preview screen's EMG tabs, and every stage is written to the diagnostic figures. The bold envelope below is the RMS the analysis actually measures: the raw one is dominated by the heartbeat (the R-waves run off the shared scale), and removing the ECG then reducing the noise drops the between-breath floor while keeping the bursts — the signal-to-noise of the inspiratory pattern roughly doubles from step 2 to step 3:

Diaphragm EMG conditioning stages

Entropy

Sample Entropy is calculated per breath for the selected channels and averaged like the other measurements, in Setup's "Sample entropy" card (shown once a channel is assigned to Entropy in the channel picker). The two settings there are named after what they literally are, not after the neighbouring textbook parameter: Template length (m + 1) is one more than the embedding dimension m used in the sample-entropy literature — the app's default of 3 gives m = 2, the value conventional there (set 2 for the m = 1 RespMech reported before this default changed). Tolerance (r), × SD is a multiple of the per-column standard deviation, not an absolute tolerance — published values are typically 0.1-0.25 × SD, with 0.2 × SD the most widely used default (Richman & Moorman, Am J Physiol Heart Circ Physiol 2000;278:H2039-49; Yentes et al., Ann Biomed Eng 2013;41:349-65); RespMech defaults to the lower end, 0.1 × SD. A read-out under the two fields states the resulting m and r in those terms, and the same wording is recorded in each output workbook's Provenance sheet whenever entropy is actually computed. Because both m and r affect the value, and r is rescaled per segment against that segment's own standard deviation, entropy values are only comparable across files and channels that share a sampling frequency (and resampling setting) and the same m and r.

When the volume column is one of the entropy columns, its sample entropy is computed on the volume RespMech itself analyses (zeroed, and drift- and trend-corrected as configured), just as an EMG column is measured on the processed EMG. If volume is integrated from flow and has no column of its own, tick Entropy on derived volume in the same card (input.channels.entropy_derived = ["volume"]) to get the same three columns. Analyses written before this rule show a notice once when they are opened, since their volume-column entropy values change.

1) Lozano-García M, Sarlabous L, Moxham J, Rafferty GF, Torres A, Jolley CJ, Jané R. Assessment of inspiratory muscle activation using surface diaphragm mechanomyography and crural diaphragm electromyography. Annu Int Conf IEEE Eng Med Biol Soc. 2018;2018:3342-3345. doi:10.1109/EMBC.2018.8513046.

2) Aboy M, Cuesta-Frau D, Austin D, Micó-Tormos P. Characterization of sample entropy in the context of biomedical signal analysis. Annu Int Conf IEEE Eng Med Biol Soc. 2007;2007:5943-5946. doi:10.1109/IEMBS.2007.4353701.


Output

Everything lands in the output folder you choose:

  • data/ — Excel workbooks: the across-file averages, and (optionally) breath-by-breath values per file, plus a cohort summary. Each workbook carries its own Units and Provenance sheets.
  • diagnostics/ — vector PDF figures per file: Campbell/PV loops (averaged and per breath), the analysed and raw signals, the staged volume correction, and per-channel EMG overviews at each conditioning stage (raw → ECG-removed → noise-reduced). Optionally the EMG channels as WAV.
  • analysis-used.toml and run-report.txt — the exact settings and a log of what was read, kept, excluded and written, so a folder of results carries its own recipe.

Use the diagnostic figures (or the Preview screen) to spot breaths to exclude — e.g. IC manoeuvres or coughs — then exclude them by clicking them in Preview.

RespMech — the Run & results drawer, expanded after a run

Correctness

The v2 engine is a port of the original v1 monolith. Golden/characterisation tests pin breath timing, volumes, ventilation, the pressure descriptors and the elastic work-of-breathing component byte-for-byte against references baked from the original implementation, which is kept frozen in legacy/ for exactly that purpose. Two deliberate deviations from 1.x, and a further SciPy-driven drift in the Simpson-integrated columns (a library version difference, not a RespMech choice), are instead pinned to within 1 part in 10⁹ (rtol 1e-9, atol 1e-12) — see Changes from RespMech 1.x.


License and usage

This program is free software: you can redistribute it and/or modify it under the terms of the GNU General Public License as published by the Free Software Foundation, either version 3 of the License, or (at your option) any later version.

This program is distributed in the hope that it will be useful, but WITHOUT ANY WARRANTY; without even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE. See the GNU General Public License for more details.

Read the entire licence here.

Sample entropy is vendored from pyEntropy — see LICENSE pyentrp.

Note to respiratory scientists

I created this code for my own work and shared it hoping that other researchers working with respiratory physiology might find it useful. If you have questions or suggestions that would make it more useful, please drop me an email.

How do I cite this code in scientific papers – and should I?

It is up to you, really. Personally I am a fan of transparency and Open Source / Open Science and I would appreciate a mention. This will also make readers of your papers aware that this code exists – if you found it useful, perhaps they will too.

Every released version has its own DOI. Reference the latest via DOI, or cite a specific version using that version's DOI (click the badge for the list). See CHANGELOG.md for what changed between versions.

An example citation:

[...] were calculated using the Python package RespMech (E Walsted, RespMech v2.5.0, 2026, https://github.com/emilwalsted/respmech/, DOI: 10.5281/zenodo.3270825) [...]

Metadata

Release files for respmech 2.5.0

For a detailed explanation of source distributions (sdists) and built distributions (wheels), please see the package formats documentation.

Source distribution (sdist)

Source distribution for respmech 2.5.0
File Size Uploaded
respmech-2.5.0.tar.gz 801.2 kB Details

Built distribution (wheel)

Table of built distributions (wheels) for respmech 2.5.0
File Interpreter ABI Platform
respmech-2.5.0-py3-none-any.whl Python 3 none any Details

Total release size: 1.7 MB

Release files / respmech-2.5.0.tar.gz

Download URL respmech-2.5.0.tar.gz
Size 801.2 kB
Tags Source
SHA-256 checksum
How to use checksums
79ec011b70d49fc13ad2b53a003eaf96b3517ffb0e09f79ca874e16981fb28dd
BLAKE2b-256 checksum
How to use checksums
de2e8b4e0efbe3ccf7be7fa0f0542f0664bebb0ac3472a3aebc24d9679ba05c0
Upload date
Uploaded using Trusted Publishing?
What is trusted publishing?
Yes
Uploaded via twine/7.0.0 CPython/3.13.14

Release files / respmech-2.5.0-py3-none-any.whl

Download URL respmech-2.5.0-py3-none-any.whl
Size 853.0 kB
Tags Python 3
SHA-256 checksum
How to use checksums
844207e16e3a282756c2649802b53f9756cf8f4785acc0f4c6f5e6a80952ed0e
BLAKE2b-256 checksum
How to use checksums
0cc2e5ac1f9156ea95cea83e3a943c9f3a164e4dd48879f58a46400c787c7a7e
Upload date
Uploaded using Trusted Publishing?
What is trusted publishing?
Yes
Uploaded via twine/7.0.0 CPython/3.13.14

Release history Release notifications | RSS feed

This release

2.5.0 This release

2 release files

2.4.0

2 release files

2.3.4

2 release files

2.3.3

2 release files

2.3.2

2 release files

2.3.1

2 release files

2.3.0

2 release files

2.2.2

2 release files

Anthropic, PBC Visionary sponsor Bloomberg Visionary sponsor Hudson River Trading Visionary sponsor Meta Visionary sponsor NVIDIA Visionary sponsor Microsoft Sustainability sponsor Depot Continuous Integration AWS Cloud computing and Security Sponsor Datadog Monitoring Fastly CDN Google Download Analytics Sentry Error logging StatusPage Status page