psgscoring

Open-source AASM-compliant respiratory scoring for polysomnography.

PyPI Python License Tests

Paper

Rombaut B, Rombaut B, Rombaut C, et al. Graded evidence in place of thresholds: an open-source, AASM-compliant method for respiratory event detection in polysomnography. Manuscript in preparation, 2026.

Technical details (signal processing chain, classification logic, bias corrections): Technical Reference (Online Supplement)

What this library does

psgscoring detects and classifies respiratory events (apneas, hypopneas, RERAs) in polysomnography recordings following AASM rules. It extends YASA (Vallat & Walker, eLife 2021) from sleep staging into a complete clinical respiratory scoring pipeline.

Four things that distinguish this library:

  1. Graded evidence instead of a threshold cascade — the AASM Rule 1A conjunction is evaluated as a product of graded terms (flow reduction, duration, desaturation-or-arousal, breath template) rather than a chain of yes/no cuts. One operating point replaces three parameter combinations. On MESA (n=150, held out) this raises event-level agreement over the rule cascade by a median ΔF1 of +0.029 (p = 7·10⁻⁸) without costing anything on the AHI itself.
  2. AHI confidence interval — every study is scored at three stringency levels (strict/standard/sensitive), yielding a per-study robustness grade (A/B/C) rather than a single AHI number.
  3. Clinical auditability — every event carries a confidence score, the rule that admitted it, the graded terms behind that decision, and the counters of every correction that touched it. Rejected candidates are kept with their rejection reason rather than discarded, so a reviewing physician can see what was not scored and why.
  4. Measured, not asserted — every behavioural change ships with the measurement that motivated it and a decision rule fixed before the sweep. CHANGELOG.md carries the numbers, including the ones that argued against the change. Validation runs on PSG-IPA (5 recordings, 12 scorers each) and MESA/NSRR (held out, nothing tuned on it).

Release policy

Releases are measurement points, not milestones. Several may be cut on the same day: every behavioural change is validated against the golden harness and at least one cohort before it ships, and the version is what makes that validation citable. A high version number reflects how often the work is measured, not instability.

The stability guarantee lives in the profiles, not in the version number:

  • mesa_shhs and chicago_1999 are frozen. They reproduce published results (paper v31/v37 and the 1999 Chicago Criteria respectively) and are pinned against behavioural changes, including repairs that every other profile receives.
  • All other profiles track the current best understanding of the AASM rules and may move between releases when a measurement justifies it. Each such change is behind a profile field whose default preserves the previous behaviour, is recorded in CHANGELOG.md with the measurement that motivated it, and is pinned by a test.

If you need scored values to stay identical across time — for a study, a regulatory submission, or a paper — pin the version (psgscoring==0.19.1) and record which profile you used. Do not rely on a profile name alone.

Installation

pip install psgscoring

Requirements: Python ≥3.9, numpy, scipy, mne. No GPU required.

Quick Start

import mne
from psgscoring import run_pneumo_analysis

# Load EDF and provide a hypnogram (e.g., from YASA)
raw = mne.io.read_raw_edf("recording.edf", preload=True)
hypnogram = ["W", "N1", "N2", "N2", "N3", ...]  # per 30-s epoch

# Run the full pipeline
results = run_pneumo_analysis(raw, hypnogram, scoring_profile="aasm_v3_rec")

# Access results
resp = results["respiratory"]["summary"]
print(f"AHI: {resp['ahi_total']}, Severity: {resp['severity']}")
print(f"Events: {resp['n_obstructive']} OA, {resp['n_hypopnea']} Hyp")

# AHI confidence interval
interval = results["ahi_interval"]
print(f"AHI interval: [{interval['strict']['ahi']}{interval['sensitive']['ahi']}]")
print(f"Robustness: {interval['robustness_grade']}")

Scoring Profiles

Parameter Strict Standard Sensitive
Hypopnea threshold ≥30% ≥30% ≥25%
SpO₂ nadir window 30 s 45 s 45 s
Peak-based detection No Yes Yes

list_profiles() enumerates the full registry — 19 profiles: AASM v1/v2/v3, CMS/Medicare, Chicago 1999, the NSRR dataset profile, and the exploratory arms (breath-graded, dual-sensor, and the four on the envelope axis below).

aasm_v3_breath — breath-graded hypopnea scoring (v0.13.0, opt-in)

run_pneumo_analysis(raw, hypnogram, scoring_profile="aasm_v3_breath")

A second hypopnea detector that takes the breath, not the sample, as the unit. The AASM speaks of peak signal excursions; this scores them directly. Five differences from the default detector:

  1. Breath-level events. An event is a run of consecutive breaths below the threshold, so boundaries land on breath transitions and a recovery breath ends the run by itself — no smoothing, and the merge/split problem that depresses event-F1 largely disappears.

  2. Two-pass patient calibration. Pass 1 finds the incontestable events; pass 2 measures the baseline from only the breathing that is not an event, and derives that patient's own SpO₂ delay by cross-correlation instead of assuming a fixed 30–45 s window.

  3. Graded AASM predicates. The rule structure stays literally the Rule 1A conjunction — ≥30% reduction AND ≥10 s AND (≥3% desaturation OR arousal) — but each threshold gets a tolerance rather than being infinitely sharp.

  4. Every event carries p_scored plus a criteria dict giving each predicate's contribution, so the audit trail says why, not just what.

    p_scored ranks events by how well they satisfy the AASM conjunction. It is not the probability that a scorer would mark the event: measured against the 12-scorer fractions on PSG-IPA (163 events) the correlation is only r = 0.194 and the level is +0.33 too high. Use it to order events, not as a likelihood.

  5. One strictness axis (hypopnea_strictness, default 0.50) instead of three parameter combinations.

Scope is hypopneas only; apneas keep the existing detector.

Status — opt-in, not the default. On PSG-IPA (5 recordings, 12 scorers) the median event-F1 rises 0.343 → 0.434, the percentile within the inter-scorer distribution p6 → p17, and mean |ΔAHI| against the scorer median falls 1.84 → 0.29. On MESA the paired advantage replicates on two disjoint held-out samples of 50 (p = 0.0069 and p = 0.0016). But the two datasets disagree on the absolute level by ~16 AHI points, and until that is explained the existing clinical output stays the default. See docs/interim_conclusie_klinisch_gebruik.md.

aasm_v3_dual — apneas on both flow sensors (v0.14.0, opt-in)

run_pneumo_analysis(raw, hypnogram, scoring_profile="aasm_v3_dual")

The AASM assigns apneas to the oronasal thermistor and hypopneas to nasal pressure. Choosing one sensor means inheriting its blind spot: nasal pressure misses mouth breathing, the thermistor is too slow for short events. This profile detects apneas on both and merges them, so neither channel can veto the other.

Each merged apnea carries corroborationboth, thermistor_only or pressure_only. Nothing is discarded on that basis. Across 1785 apnea detections on MESA only 19% are seen by both sensors, so requiring corroboration would remove four events in five. Discarding can be licensed explicitly (corroborate_apnea_events(..., corroboration_licensed=True)); no profile does so.

meta["flow_channels"]["thermistor_check"] reports the envelope correlation between the two channels either way. Under this profile it is informational — the second sensor is additive and can only add events. It gates only the substitutive path, where a dead thermistor would delete them.

Why this exists. v0.13.2 made the thermistor the apnea sensor outright. On a montage where that channel carried no usable breathing signal, apneas went from 93 to 0 and the conclusion moved from moderate CSAS — human-confirmed — to mild SAS. That release was rolled back. aasm_v3_rec remains the default and is byte-identical.

aasm_v3_pressure — nasal-pressure reference (v0.14.1, opt-in)

run_pneumo_analysis(raw, hypnogram, scoring_profile="aasm_v3_pressure")

Apnea detection is not the only consumer of a flow signal. The AHI robustness sweep, baseline anchoring, the arousal/RERA coupling, Cheyne-Stokes detection and the ventilatory burden each take one — and each took the apnea channel. This profile scores apneas exactly as aasm_v3_rec does and points those five at the nasal pressure instead.

profile apneas derived analyses
aasm_v3_rec thermistor¹ thermistor¹
aasm_v3_pressure thermistor¹ nasal pressure
aasm_v3_dual both, merged nasal pressure

¹ where it passes the quality check; nasal pressure otherwise.

The AASM assigns quantitative flow assessment to nasal pressure and treats the oronasal thermistor as qualitative — able to show that flow stopped, not to grade how much of it is left. Reading a thermistor for the ventilatory burden therefore measures something the sensor does not claim to report: on one recording it read 20.4% instead of 42.6%, below the ≤25% reference, for a patient spending 94.7% of the night under 90% saturation.

Identical to aasm_v3_rec on any montage without a usable thermistor. Set flow_reference on any profile to get the same behaviour; meta["flow_channels"]["reference_sensor"] reports which channel was used.

The envelope axis — four exploratory arms (v0.19.0, all off)

Every threshold in the library is applied to an amplitude envelope, and until now there was only one way to build it: the analytic signal over the whole recording. That transform is also the library's largest allocation. On a 12 h recording at 256 Hz, one preprocess_flow call costs 592 MB of peak RSS and 34 s, which is what makes a parallel sweep swap.

envelope_method and envelope_fs make the choice explicit. Measured on mesa-sleep-0001.edf with /usr/bin/time -v, peak above a 701 MB load-only floor:

profile envelope_method peak time what changes
aasm_v3_rec hilbert 592 MB 33.7 s reference — every published result
aasm_v3_env_chunked hilbert_chunked 116 MB 0.8 s same transform, 30-min blocks
aasm_v3_env_rectify rectify_lowpass 255 MB 0.9 s AM demodulation
aasm_v3_env_breath breath_amplitude 92 MB 1.1 s per-breath amplitude, interpolated
aasm_v3_env_decimated hilbert, envelope_fs=10 152 MB 1.1 s decimate, transform, interpolate back

None of these is a default, and the blockwise one is not free. It was planned as an implementation detail — same numbers, no profile field — on the reasoning that a generous overlap reproduces the full transform up to numerical noise. It does not: the 1/(πt) kernel has no compact support, so the interior residual floors around 1e-4 of the p95 envelope and does not converge as the overlap grows, while the first and last samples differ by ~30 % because an FFT wraps circularly. Far below any clinical threshold, but the envelope is compared to a baseline sample by sample, so it moves boundaries.

The golden harness cannot referee this axis — its cases are shorter than one block, so chunking never engages and every case passes regardless. The tests that do use 8 h signals; see tests/test_envelope_methods.py.

Measured against human scoring on PSG-IPA (n = 5, twelve scorers each, anchor in the same run — bias +1.69/h, MAE 1.76/h, F1 0.462):

arm F1 bias MAE verdict
aasm_v3_env_chunked 0.462 +1.69 1.76 identical to the anchor; 116 of 8350 boundaries move, each by one 0.1 s step
aasm_v3_env_rectify 0.487 +0.01 1.72 better on nearly every axis — and still off, see below
aasm_v3_env_breath 0.414 +3.19 4.11 fails; events lost to the IoU threshold go 63 → 205
aasm_v3_env_decimated 0.463 +1.71 1.78 within noise, one extra event

And on MESA (n = 150, held out, paired per recording, anchor F1 0.438 / bias −5.30 in the same job):

arm F1 med bias paired ΔF1 p verdict
aasm_v3_env_chunked 0.436 −5.29 +0.0000 0.70 equal on 129/150 recordings
aasm_v3_env_rectify 0.416 −5.67 −0.0015 0.10 the PSG-IPA gain did not replicate
aasm_v3_env_breath 0.413 −5.78 −0.0206 2.4·10⁻⁶ rejected on both cohorts
aasm_v3_env_decimated 0.438 −5.30 +0.0000 0.91 equal on 111/150 recordings

rectify_lowpass is why the decision rule was fixed before the measurement. It was the best arm on PSG-IPA on nearly every axis; on 150 held-out recordings the sign flips and the advantage is gone. Five recordings had described those five recordings. Had the rule been written afterwards, it would have shipped.

hilbert_chunked now meets its promotion criterion on both cohorts — same bias, same mean F1, identical severity confusion matrix — at 592 MB → 116 MB and 33.7 s → 0.8 s. It is still not promoted: that is a decision about the published numbers (the MESA F1 median moves 0.438 → 0.436), not about the code. The rule and the full numbers, including boundary offsets against the human-versus-human distribution, are in the CHANGELOG.

Validation

PSG-IPA (PhysioNet): 5 recordings, 59 independent scorer sessions. Against the scorer median, aasm_v3_rec gives bias +1.69/h, MAE 1.76/h, Pearson r = 0.997, weighted κ = 0.839, severity concordance 4/5. Note the sample size: five recordings is enough to expose a defect, not enough to estimate a population.

MESA (NSRR, external cohort, n = 150): nothing is tuned on it. Against a reconstructed AASM-2015 reference, per profile:

profile median F1 precision recall bias (/h) MAE (/h)
aasm_v3_rec 0.438 0.519 0.408 −5.30 10.12
aasm_v3_breath 0.510 0.628 0.472 −5.18 9.51
aasm_v3_prob 0.513 0.670 0.462 −6.41 9.97
aasm_v3_dual 0.446 0.498 0.490 −1.17 9.24
aasm_v3_breath_dual 0.504 0.577 0.511 −2.34 9.37

Paired against aasm_v3_rec, breath-graded scoring raises event agreement by a median ΔF1 of +0.029 (95/150 recordings, p = 6.8·10⁻⁸); the probabilistic variant by +0.036 (p = 1.4·10⁻⁹).

These numbers moved substantially in 0.17.0, and the reason is worth stating plainly. The earlier figures on this cohort reported a bias of −11 to −15/h and were read as under-detection. Most of it was not: a signal-quality gate was rejecting effort channels by how the EDF declared its unit rather than by the signal in them, and it failed 52 of 52 MESA recordings. Repairing it halved the bias at identical F1, precision and recall on three of the five profiles — the events had been found all along and were being discarded in the accounting. See CHANGELOG.md 0.16.0 and docs/.

Bias corrections

The twelve below were the original systematic pass: six over-counting and six under-counting mechanisms, each identified and quantified separately. The list has grown since and is no longer the whole story — CHANGELOG.md carries the current state, with the measurement behind every entry. The most consequential recent one: a sensor-quality gate that judged how the EDF declared its unit rather than the sensor itself.

# Correction Direction Clinical impact
1 Post-apnea baseline inflation Over-counting Prevents false Mild→Moderate
2 SpO₂ cross-contamination Over-counting Flags uncertain coupling
3 Cheyne-Stokes trough scoring Over-counting Prevents HF misdiagnosis as OSA
4 Low-confidence defaults Over-counting Confidence stratification
5 Artefact-flank exclusion Over-counting Prevents post-disconnect events
6 Local baseline validation Over-counting Rejects inflated-baseline FPs
7 Peak-based amplitude detection Under-counting AASM-conformant breath-level
8 Extended SpO₂ nadir window Under-counting Catches delayed desaturations
9 Flow smoothing removal Under-counting Eliminated +54 FPs on PSG-IPA
10 Position signal auto-mapping Under-counting Handles raw ADC encoding
11 Configurable profiles Under-counting Sensitivity adjustment per study
12 Flattening-based RERA Under-counting Flow limitation without amplitude drop

Since then (see CHANGELOG.md for the measurements):

Correction Direction What it was
Square-root linearisation applied per profile, not per montage Over-counting The AASM Rule 3 linearisation ran only when the hypopnoea and apnoea channels differed, so the same profile pre-processed one-channel and two-channel montages differently
Scale-free RIP quality gate Under-counting An absolute amplitude threshold rejected effort channels by how the EDF declared its unit; on MESA it failed 52 of 52 recordings, halving the reported AHI bias once repaired
Stability filter across all hypopnoea subtypes Over-counting The filter compared the event type exactly against "hypopnea", so every subtyped hypopnoea escaped it
Desaturation re-use limit Over-counting One desaturation could confirm any number of adjacent events (opt-in, off by default)
Breath-granular event boundaries Neither The rule cascade starts hypopnoeas 2–5 s before human scorers; snapping to breath edges removes that offset (opt-in, off by default)
Coherence-based thermistor gate Neither The gate deciding which sensor carries apnoeas correlated amplitude envelopes, while the question is whether both sensors see the same breaths — a timing question (opt-in, off by default)
Long-event splitting Neither Splits events beyond a duration ceiling (opt-in, off by default)
Low-baseline desaturation relaxation Neither Relaxes the 3% desaturation requirement when the local baseline is already hypoxic (opt-in, off by default)

Four of these are off, and that is a result, not an omission. Each was implemented, measured, and left disabled because the measurement did not support enabling it: the desaturation limit removed 8 events in 4303; breath boundaries bought recall no precision; long-event splitting duplicated an existing splitter and found no hypopnoea candidates; the coherence gate is a real repair of a real defect but costs 2.95/h of AHI bias on MESA when switched on. They ship as switches so the finding stays reproducible and the next cohort can overturn it. docs/third_party_comparison.md records one row per idea evaluated, including the ones rejected.

Architecture

~15,000 lines across 21 submodules, 868 unit tests (CI: Python 3.9–3.12):

constants · utils · signal · breath · breath_scoring · classify · spo2 · plm · ancillary · arousal · respiratory · indices · ventilation · pipeline · ml_classifier · profiles · postprocess · signal_quality · signal_quality_channels · ecg_effort · _types

Behaviour is pinned by a golden-output harness (PSGSCORING_GOLDEN=1) that scores fixed synthetic recordings and compares a digest. Every release runs it before publishing; a difference is a stop, not a re-bless.

Related

  • YASAFlaskified — web platform integrating psgscoring with YASA staging, multilingual PDF reports, EDF+ export, and FHIR R4
  • YASA — AI-based sleep staging (Vallat & Walker, eLife 2021)
  • slaapkliniek.be — live instance (no installation required)

Citation

@article{rombaut2026psgscoring,
  title     = {Graded evidence in place of thresholds: an open-source,
               {AASM}-compliant method for respiratory event detection in
               polysomnography},
  author    = {Rombaut, Bart and Rombaut, Briek and Rombaut, Cedric},
  year      = {2026},
  note      = {Manuscript in preparation}
}

Disclaimer

psgscoring is research software — not a medical device. It is not CE-marked (MDR 2017/745) or FDA-cleared. All outputs are research-grade estimates that must be reviewed by a qualified clinician before any diagnostic or therapeutic decision. See DISCLAIMER.md for the full text.

License

BSD-3-Clause. See LICENSE.


Contact: bart.rombaut@gmail.com

Download files

Download the file for your platform. If you're not sure which to choose, learn more about installing packages.

Source Distribution

psgscoring-0.23.0.tar.gz (1.5 MB view details)

Uploaded Source

Built Distribution

If you're not sure about the file name format, learn more about wheel file names.

psgscoring-0.23.0-py3-none-any.whl (1.4 MB view details)

Uploaded Python 3

File details

Details for the file psgscoring-0.23.0.tar.gz.

File metadata

  • Download URL: psgscoring-0.23.0.tar.gz
  • Upload date:
  • Size: 1.5 MB
  • Tags: Source
  • Uploaded using Trusted Publishing? Yes
  • Uploaded via: twine/7.0.0 CPython/3.13.14

File hashes

Hashes for psgscoring-0.23.0.tar.gz
Algorithm Hash digest
SHA256 ac31aba572cfdb52e8ba6ce0bb4bd4e5c8018ad70ee0491f81a86027c81768b5
MD5 018be2378cc617d0a3baa59de1d93097
BLAKE2b-256 afbf34ef599f49b37667543ceecd450ac3c1d96067be9321c9b6f1f71caac637

See more details on using hashes here.

Provenance

The following attestation bundles were made for psgscoring-0.23.0.tar.gz:

Publisher: publish.yml on bartromb/psgscoring

Attestations: Values shown here reflect the state when the release was signed and may no longer be current.

File details

Details for the file psgscoring-0.23.0-py3-none-any.whl.

File metadata

  • Download URL: psgscoring-0.23.0-py3-none-any.whl
  • Upload date:
  • Size: 1.4 MB
  • Tags: Python 3
  • Uploaded using Trusted Publishing? Yes
  • Uploaded via: twine/7.0.0 CPython/3.13.14

File hashes

Hashes for psgscoring-0.23.0-py3-none-any.whl
Algorithm Hash digest
SHA256 a15aba1b242d91df299201e8ca13e9a98a0a3bfb554bdd2dc5e44fd2bc370476
MD5 9852a9bdd16e51533c37a4876de62fc4
BLAKE2b-256 52094beed3ae2b42069f67438957a41742c91742eb36045856d3dfff430831a0

See more details on using hashes here.

Provenance

The following attestation bundles were made for psgscoring-0.23.0-py3-none-any.whl:

Publisher: publish.yml on bartromb/psgscoring

Attestations: Values shown here reflect the state when the release was signed and may no longer be current.

Release history Release notifications | RSS feed

0.32.0

2 files

0.31.6

2 files

0.31.5

2 files

0.31.4

2 files

0.31.3

2 files

0.31.2

2 files

0.31.1

2 files

0.31.0

2 files

0.30.0

2 files

0.29.0

2 files

0.28.1

2 files

0.28.0

2 files

0.27.7

2 files

0.27.6

2 files

0.27.5

2 files

0.27.4

2 files

0.27.3

2 files

0.27.0

2 files

0.26.0

2 files

0.25.0

2 files

0.24.0

2 files

This release

0.23.0 This release

2 files

0.22.0

2 files

0.21.0

2 files

0.20.0

2 files

0.19.1

2 files

0.19.0

2 files

0.18.0

2 files

0.17.0

2 files

0.15.2

2 files

0.15.1

2 files

0.15.0

2 files

0.14.9

2 files

0.14.8

2 files

0.14.7

2 files

0.14.6

2 files

0.14.5

2 files

0.14.4

2 files

0.14.3

2 files

0.14.2

2 files

0.14.1

2 files

0.14.0

2 files

0.13.2

2 files

0.13.1

2 files

0.13.0

2 files

0.12.2

2 files

0.12.1

2 files

0.12.0

2 files

0.11.0

2 files

0.10.0

2 files

0.9.0

2 files

0.7.6

2 files

0.7.3

2 files

0.7.2

2 files

0.7.1

2 files

0.7.0

2 files

0.6.2

2 files

0.6.1

2 files

0.6.0

2 files

0.5.2

2 files

0.5.1

2 files

0.5.0

2 files

0.4.5

2 files

0.4.4

2 files

0.4.3

2 files

0.4.2

2 files

0.3.2

2 files

0.3.1

2 files

0.2.963

2 files

0.2.962

2 files

0.2.961

2 files

0.2.951

2 files

0.2.96

2 files

0.2.95

2 files

0.2.94

2 files

0.2.93

2 files

0.2.92

2 files

0.2.91

2 files

0.2.9

2 files

0.2.8

2 files

0.2.7

2 files

0.2.6

2 files

0.2.4

2 files

0.2.3

2 files

0.2.2

2 files

0.2.0

2 files

Anthropic, PBC Visionary sponsor Bloomberg Visionary sponsor Hudson River Trading Visionary sponsor Meta Visionary sponsor NVIDIA Visionary sponsor Microsoft Sustainability sponsor Depot Continuous Integration AWS Cloud computing and Security Sponsor Datadog Monitoring Fastly CDN Google Download Analytics Sentry Error logging StatusPage Status page