TemField
Radiated EMC susceptibility measurements in (G)TEM waveguides using mpylab
This software is distributed unter GPL-3 or higher. See LICENSE for details.
Installation
pip install temfield-mpylab
For an isolated application-style installation, pipx is also useful:
pipx install temfield-mpylab
This installs the package in its own environment and exposes the command line
tools temfield and temfield-info on your PATH.
Command line
Start the GUI with:
temfield
Show installation and dependency information with:
temfield-info
Use temfield-info --json for machine-readable output.
AM headroom safety check
Before modulation is enabled, TEMField first levels the unmodulated field to
the AM peak factor. For 80 % AM this is 1.8 * E_target. It then reduces the
signal-generator output by 5.1 dB and checks the resulting decrease. A
configured forward-power meter provides the measurement method prescribed by
IEC 61000-4-20; otherwise the field probe is used as an explicitly
non-normative proxy. After a passed check, TEMField levels once more to the
requested carrier field and only then enables AM. Failed leveling or headroom
checks keep AM off and switch RF off for that frequency.
By default mpylab applies the nominal interval 3.10545...7.1 dB for 80 % AM. IEC 61000-4-3:2020 explicitly uses the 3.1...7.1 dB interval. IEC 61000-4-20:2022 retains older wording that declares 3.1...5.1 dB suitable and values below 3.1 dB saturated, but does not consistently resolve values above 5.1 dB. TEMField deliberately applies the current IEC 61000-4-3 linearity criterion to its IEC 61000-4-20 workflow as a documented engineering interpretation. The withdrawn IEC 61000-4-3:2006/ISH1:2008 supports accepting reductions above 5.1 dB but is not presented as a current normative reference. The corresponding IEC publication records are ISH1:2008, IEC 61000-4-3:2020, and IEC 61000-4-20:2022.
Both interval limits use nominal values: uncertainty does not change pass/fail
by default. The mpylab API makes the upper limit configurable and can
optionally allow expanded-uncertainty overlap only at that upper limit;
TEMField leaves this option disabled. Programmatic integrations can set
am_headroom_maximum_drop_db and
am_headroom_allow_upper_uncertainty on TEMFieldWorker or
TestSusceptibility.Init. Ratio uncertainty is retained with coverage factor
k = 2; shared calibration contributions remain correlated and can cancel in
the ratio, while independent acquisition scatter remains.
The log and JSON result distinguish passed, passed_with_uncertainty,
failed_below_interval, and failed_above_interval. They record both nominal
limits, the uncertainty option, standard and expanded asymmetric uncertainties,
decision interval, method, and normative-method flag.
Field-probe coordinate contract
mpylab field-probe drivers return raw (probe_x, probe_y, probe_z) values.
TEMField resolves the probe orientation once and maps those readings into
(cell_x, cell_y, cell_z). The GUI selections, datafunc, leveling,
AM-headroom field proxy, waveform plot, sinusoidal fit, and result files all
operate only on this mapped cell-coordinate vector.
Define the physical probe orientation on its DOT node, for example:
prb [ini="fieldprobe.ini"
probe_axis_map="cell_x:+probe_y,cell_y:+probe_x,cell_z:-probe_z"]
The same definition may be stored in the DESCRIPTION or Init_Value
section of a FIELDPROBE INI file. Programmatic integrations may pass
probe_orientations to TEMFieldWorker or TestSusceptibility.Init.
Conflicting definitions stop the run before hardware initialization.
The default leveling component is Ey. TEMField remembers the last selected
component. Installations with the removed legacy auto setting are migrated
to Ey when their settings are next loaded.
Custom TestSusceptibility.Init(datafunc=...) functions receive the mapped
three-component cell-coordinate vector. Current probes provide unsigned
component magnitudes, so TEMField supports axis permutations but not arbitrary
rotations that require synchronous signed vector data.
EUT monitoring
TEMField uses the shared mpylab.env.eut API. Manual EUT reporting is
always available in the GUI and may be supplemented by automatic monitors for
cameras, communication links, or process data. Automatic monitors run behind
ThreadedEUTMonitor; they never access MGraph or switch RF themselves.
During exposure, the operator can report Degraded, Failed, or Not evaluable. After an impairment, TEMField switches AM and RF off and separately
records the post-exposure state and recovery method. The table reports the
result for performance criterion A, B, or C and includes the structured events
in its tooltip. The post-exposure timeout is configurable in the EUT panel. A
safety event is routed immediately through the measurement worker's RF-off
path.
RandomEUTMonitor provides reproducible hardware-independent simulations and
a starting point for custom automatic monitors. It must not be used as evidence
of real EUT performance.
An application embedding the window can configure an automatic monitor before the measurement starts:
from mpylab.env.eut import RandomEUTMonitor
window.set_automatic_eut_monitor(RandomEUTMonitor(seed=42))
Structured results
Saving the measurement table writes both the selected CSV file and a JSON file
with the same stem. The JSON document preserves quantities as value,
uncertainty, and unit and includes the EUT event history and performance
assessment.
The shared mpylab.env.immunity_result schema calls the applied test quantity
disturbance. Its
quantity_kind, target, and freely named measured_components are not tied
to an electric field. TEMField currently records electric_field_strength
with its measured probe components. The same format can represent, for
example, injected_current in A or an applied voltage in a future conducted
immunity application.
License
GPL-3 or higher
Repository
Documentation
https://temfield-1012e2.gp.hrz.tu-chemnitz.de/
Contact
Prof. Dr. Hans Georg Krauthäuser (hgk@ieee.org)
Chair for Electromagnetic Theory and Compatibility
Technische Universität Dresden, Dresden, Germany
Metadata
Release files for temfield-mpylab 0.0.12
For a detailed explanation of source distributions (sdists) and built distributions (wheels), please see the package formats documentation.
Source distribution (sdist)
| File | Size | Uploaded | Provenance | Actions |
|---|---|---|---|---|
| temfield_mpylab-0.0.12.tar.gz | 80.6 kB | Without provenance | Details for temfield_mpylab-0.0.12.tar.gz |
Built distribution (wheel)
| File | Interpreter | ABI | Platform | Actions Reset |
|---|---|---|---|---|
| temfield_mpylab-0.0.12-py3-none-any.whl | Python 3 | none | any | Details for temfield_mpylab-0.0.12-py3-none-any.whl |
Total release size: 119.5 kB
Release files / temfield_mpylab-0.0.12.tar.gz
| Download URL | temfield_mpylab-0.0.12.tar.gz |
|---|---|
| Size | 80.6 kB |
| Tags | Source |
|
SHA-256 checksum How to use checksums |
5aaa4c92d53007de839192c112f9392d9c799911f3b0724f5368a1ca849dffaa
|
|
BLAKE2b-256 checksum How to use checksums |
8c79f2aa91ea93eff21d108a5207f2954daad98a3c7c5d08427461a3b1893dec
|
| Upload date | |
|
Uploaded using Trusted Publishing? What is trusted publishing? |
No |
| Uploaded via |
twine/7.0.0 CPython/3.13.5
|
Release files / temfield_mpylab-0.0.12-py3-none-any.whl
| Download URL | temfield_mpylab-0.0.12-py3-none-any.whl |
|---|---|
| Size | 39.0 kB |
| Tags | Python 3 |
|
SHA-256 checksum How to use checksums |
d9f37132f6808bf5466811d1b2ec8c35b2ccd755688f8edd3b880d944658d866
|
|
BLAKE2b-256 checksum How to use checksums |
41625ef5316435bb6452a6138610f32dfa1535f9517a6174e20c3d2aca999026
|
| Upload date | |
|
Uploaded using Trusted Publishing? What is trusted publishing? |
No |
| Uploaded via |
twine/7.0.0 CPython/3.13.5
|
Security for temfield-mpylab 0.0.12
Provenance
Provenance describes where a file came from. On PyPI, provenance is shared via attestations, which provide a verifiable record of the build or publishing details. View details, limitations and caveats.
No provenance found
None of the files in this release were uploaded with provenance metadata.
A note on package security
PyPI is a repository for distributing packages, not a service for auditing them. While third-party security researchers and our staff do their best to detect and remove malicious packages, PyPI cannot guarantee the quality, security or suitability of any given package.
To learn more about protecting yourself from malicious actors, visit our package security documentation.